摘要 根据国家标准GB/T11906-1989《高碘酸钾氧化光度法测定锰[Ⅱ]》的原理,建立了水中锰[Ⅱ]的流动注射全自动在线分析法,并对各种实验条件进行了优化。实验结果表明:该方法快速、灵敏、试剂消耗量少,线性范围为0.01~10mg/L,最低检出限为0.017mg/L,相对标准偏差为1.21%(n=12)。该方法能够很好地用于地表水、地下水及工业废水中锰[Ⅱ]的在线自动监测。 A flow injection spectrophotometric analysis method for on-line determination of Mn2+ in water is proposed,and an automatic on-line Mn2+ in water analyzer was developed based on the principle of potassium periodate spectrophotometry.The experimental conditions have been optimized.The experimental results indicate that the method has higher anti-interference ability,and the Mn2+can be determined in the range of 0.01~10mg/L with the detection limit of 0.017mg/L.The RSD is 1.21% (n=12).The method can be applied to the on-line determination of Mn2+ in surface water, ground water,and industrial waste water.
出处 《分析仪器》 CAS 2009年第6期54-57,共4页 Analytical Instrumentation
  • 相关文献

参考文献3

  • 1魏复盛.水和废水监测分析指南(第四版).北京:中国环境科学出版社,1994:338.
  • 2国家环境保护总局.水和废水监测分析方法(第3版)[M].北京:中国环境科学出版社,1989,5..
  • 3Yamane T, Koshino K. Japan Talanta, 1996, 43: 963- 969.