全自动尿沉渣分析仪UF-100中细菌计数参数阈值的研究 被引量:1

Evaluation of Threshold of Parameter on Bacterial Particle of Automated Urine Flow Cytometry UF-100
原文传递
导出
摘要 目的:研究尿分析仪UF-100中细菌计数参数的阈值。方法:以462份住院患者清洁中段尿液为标本,以细菌定量培养为对照,建立ROC曲线确定阈值;通过拟合曲线确定细菌计数和定量培养结果相互关系。结果:以培养结果102CFU/ml、104CFU/ml和105CFU/ml为阈值,细菌计数阈值分别是5459/μl、6520/μl和8092/μl。不同性别、不同菌种阈值有不同。培养计数对数值(Y)-细菌计数值(X)的直线回归是Y=1.9302+0.0002X,对数回归是Y=-6.1412+1.1214×lnX。结论:定量培养确定UF-100中细菌计数参数的阈值,该值随性别、菌种不同而异。通过回归曲线及细菌计数结果可以预测定量培养结果。 Objective:To evaluate the threshold of parameter on bacterial particle of UF-100. Method:Urine quantitative cultures of 46Z samples from hospitalized patients were used as a reference method for comparison. Threshold of parameter on bacterial particle was determined by ROC curve. The association between the quantity of bacterial particle measured by UF-100 and the quantity of urine cultures was demostrated by regression curve. Results:The threshold of bacterial particle were 5 459/μl,6 520μl and 8 092/μl compared with culture at a cutoff of 10^2CFU/ml, 104CFU/ml and 10s CFU/ml,respectively. The threshold of bacterial particle were influenced by gender and species of pathogen. The regression curves of the quantity(X) of bacterial particle and the logarithm(Y) of the quantity of cultures were Y: 1. 930 2+0. 000 2X and Y=-6. 141 2+1. 121 4×1nX. Conclusions:The threshold of bacterial particle is determined by urine quantitative culture and influenced by gender and species of pathogen. The results of urine culture can be estimated by the regression curve.
出处 《中国误诊学杂志》 CAS 2009年第36期8824-8826,共3页 Chinese Journal of Misdiagnostics
关键词 尿分析/仪器和设备 集落计数 微生物 Urinalysis/instrumentation Colony Count Microbial
  • 相关文献

参考文献11

  • 1U. S. Preventive Services Task Force. Screening for Asymptomatic Bacteriuria in Adults:U. S. Preventive Services Task Force Reaffirmation Recommendation Statement [J]. Ann Intern Med, 2008,149:43-47.
  • 2Bent S, Nallamothu BK,Simel DL,et al. Does this woman have an acute uncomplicated urinary tract infection[J]. JAMA, 2002,287 (20) :2701-2710.
  • 3Mayo S,Acevedo D,Quinones-Torrelo C,et al. Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments[J]. J Clin Lab Anal, 2008,22 (4):262-270.
  • 4Ito K,Tatsumi N, Kikuchi H, et al. Basic evaluation of the new fully automated urine sediment analyzer (U-FCM) based on flowcytometric technology[J]. Rinsho Byori, 1994,42 (10): 1093- 1098.
  • 5KimSY,KimYJ,Lee SM,et al. Evaluation of the SysmexUF 100 urine cell analyzer as a screening test to reduce the need for urine cultures for community-acquired urinary tract infection[J].Am J Clin Pathol,2007,128(6):922 925.
  • 6dos Santos JC,Weber LP,Perez LR. Evaluation of urinalysis parameters to predict urinary-tract infection[J]. Braz J Infect Dis, 2007,11 (5) :479-481.
  • 7Koken T, Aktepe OC, Serteser M, et al. Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF- 100) in urinary tract infections[J]. Int Urol Nephrol, 2002,34 (2):175-178.
  • 8Evans R, Davidson MM,Sim LR, et al. Testing by Sysmex UF- 100 flow cytometer and with bacterial culture in a diagnostic laboratory : a eomparison[J]. J Clin Pathol, 2006,59 (6) :661-662.
  • 9Fenili D,Pirovano B. The automation of sediment urinalysis using a new urine flow cytometer(UF-100)[J]. Clin Chem Lab Med, 1998,36(12) :909-917.
  • 10Zaman Z, Roggeman S,Verhaegen J. Unsatisfactory performance of flow cytometer UF 100 and urine strips in predicting outcome of urine cultures[J]. J Clin Microbiol, 2001,39(11) : 4169-4171.

二级参考文献15

  • 1林萍,孙阳,张友玲,方心驰,李泳.UF-100尿沉渣分析仪及干化学分析在筛检尿路感染中的应用[J].检验医学,2004,19(4):307-309. 被引量:19
  • 2陈卫宾,李莉雅,高锋,庄亦辉,王坚镪.iQ200全自动尿液显微镜分析仪筛检尿路感染[J].检验医学,2006,21(5):505-508. 被引量:13
  • 3Hiroshi O, Toshiri S, Akinobu G,et al. Enumeration of bacterial cell Numbers and Detection of Significant Bacteriuria by Use of a New Flow Cytometry-Based Device[J]. Clin Micro, 2006, 44 (10) : 3596-3599.
  • 4Evans R, Davidson MM, Sim LR, et al. Testing by Sysmex UF- 100 flow cytometer and with bacterial culture in a diagnostic labo ratory:a comparison [J]. Clinical Pathology, 2006, 59 (6): 661-662.
  • 5Hiroshi okada, Yutaka sakai, Shigenori Miyazaki, et al. Detection of significant bacteriuria by automated urinalysis using flow cytometry[J]. Clin Micro, 2000,38 (8) : 2870-2872.
  • 6Koken T,Aktepe C,SerteserM ,et al. Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF-100) in urinary tract infections [J]. Int Urol Nephrol, 2002,34 (2) : 175- 178.
  • 7Zaman Z, Roggeman S,Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures[J]. Clin Microbiol, 2001,39 ( 11 ) : 4169-4171.
  • 8Manoni F,Valverde S,Antico F,et al. Field evaluation of a second-generation cytometer UF-100 in diagnosis of acute urinary tract infections in adult patients[J]. Clinical Microbiology Infection,2002,8(10) :662-668.
  • 9Regeniter A, Haenni V, Risch L ,et al. Urine analysis performed by flow eytometry: reference range determination and comparison to morphological findings, dipstick chemistry and bacterial culture results-a multicenter study[J]. Clinical Nephrology, 2001,55 (5) 384-392.
  • 10Juliana CS,Liliana PW,Leandro RR. Evaluation of urinalysis parameters to predict urinary-tract infection[J]. Braz J Infect Dis, 2007,11 (5):479-481.

共引文献7

同被引文献10

  • 1马筱玲,李桂琴.临床微生物学尿培养操作规范[J].中华检验医学杂志,2005,28(10):1085-1087. 被引量:38
  • 2Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures[J]. J Clin Microbiol, 2001, 39(11): 4169-4171.
  • 3Kouri TT, K:lk6nen U, Malminiemi K, et al. Evaluation of sysmex UF-100 urine flow cytometer vs chamber counting of supravitally stained specimens and conventional bacterial cultures[J]. Am J Clin Pathol, 1999, 112(1): 25-35.
  • 4Koken T, Aktepe OC, Serteser M, et al. Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF-100) in urinary tract infections[J]. Int Urol Nephrol, 2002, 34(2): 175-178.
  • 5Evans R, Davidson MM, Sire LR, et al. Testing by sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory: a comparison[J]. J Clin Pathol, 2006, 59(6): 661-662.
  • 6Fenili D, Pirovano B. The automation of sediment urinalysis using a new urine flow cytometer (UF-100)[J]. Clin Chem Lab Med, 1998, 36(12): 909-917.
  • 7Rosa RD, Grosso S, Bruschetta G, et al. Evaluation of the sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection[J]. Clin Chim Acta, 2010, 411(15-16): 1137-1142.
  • 8魏梦岚,陈伟,高洁.UF-100尿沉渣分析仪在尿路感染早期诊断中的应用[J].现代医药卫生,2008,24(4):483-484. 被引量:3
  • 9魏宏莲,范世英,王喜栋,吴亚.UF-100尿液分析仪在尿路感染筛检中的应用[J].西部医学,2011,23(2):364-365. 被引量:5
  • 10黎国庆,郭明好,杨磊,刘云.膀胱穿刺尿同步细菌培养在女性尿路感染诊断中的价值[J].郑州大学学报(医学版),2002,37(2):224-225. 被引量:5

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部