数字X线摄影系统在新生儿检查中的应用评价

Assessment of Digital Radiography System in Neonatal Diagnosis
下载PDF
导出
摘要 目的评价DR摄影系统在新生儿检查中的应用价值。方法利用CR系统和DR系统(使用剂量为前者的1/4)分别得到新生儿30幅图像。由6位医生来评价以上两者的影像系统对于导尿管末端的显影的清晰程度。结果二者的图像在诊断质量上没有统计学差异。结论DR有较好的分辨率和低噪声特性,高DQE有助于减少新生儿患者的接受剂量,图像传输速度快,使得临床诊断更为及时。 Objective To evaluate the application value of DR in neonatal diagnosis. Methods The thirty neonatal images were exposed by CR and DR(25% dosage relative to CR) respectively . The six doctors assessed the clearness degree of catheter bottom of CR and DR. Results The diagnosis quality of CR and DR did not show duce the statistical significance. Conclusion DR had better resolution and lower noise, High DQE helped to redosage of neonates, Quick transmission speed of images was propitious to clinical diagnosis quickly.
出处 《生物医学工程学进展》 CAS 2009年第4期210-212,共3页 Progress in Biomedical Engineering
关键词 数字化摄影 新生儿 剂量 Digital radiography, Neonate, Dosage
  • 相关文献

参考文献4

  • 1E. Samei, M. J. Flynn. An experimental comparison of detector performance for computed radiography systems [ J ]. Medical Physics, 2002,29 (4) :447 - 459.
  • 2E. Samei, M. J. Flynn. An experimental comparison of detector performance for direct and indirect digital radiography systems[J]. Medical Physics,2003.30(4) :608 -622.
  • 3E. Samei, Jeanne G. Hill, G. Donald, et al. Evaluation of a fiat panel digital radiographic system for low - dose portable imaging of neonates [ J ]. Medical Physics, 2003.30 (4) :604 - 607.
  • 4Klaus Bacher, Peter Smeets, Ludo Vereecken, et al. Image Quahty and Radiation Dose on Digital Chest Imaging:Comparison of Amorphous Silicon and Amorphous Selenium Flat - Panel Systems [ J ]. 2006,187 (9) :630 - 637.