摘要 采用化学抑制-电导-离子色谱法测定工业循环水中阴离子,以3.5mmol/L Na2CO3、1mmol/L NaHCO3为淋洗液。18.2mol/L H2SO4为再生液,流速1.2ml/min,5种阴离子在10.13min完成检测。各离子的检出限0.01~0.5mg/L,本法精密度小于2%,线性范围较宽,水样中各阴离子的平均加标回收率92.63%~103.57%。该法适合工业循环水中阴离子含量测定。 The chemically suppressed conductivity detection and ion chromatography were used to detect the content of five anions in industrial recycled water. They were separated simultaneously with a mixed solution of 3.5mmol/L Na2CO3 and lmmol/L NaHCO3 as absorbing solution, and 18.2 mol/L H2SO4 as regenerate solution. The samples were analyzed at the flow rate of 1.2mL/min, and were detected directly by conductimetric detector in 10. 13 minutes. Under the optimized condition,the detection limits for five analytes were from 0.01 to 0.5mg/L, RSD of the results were less than 2% ,The linear ranges were wide, average recoveries for five inorganic anions were between 92.63% - 103.57%. The method was suitable to determine the content of inorganic anions in industrial recycled water.
作者 童玲 李月霞
出处 《山东化工》 CAS 2009年第12期43-45,共3页 Shandong Chemical Industry
关键词 离子色谱 工业循环水 阴离子 含量 ion chromatography industrial recycled water inorganic anion content