两种方法对LASIK手术前后角膜厚度测量的对比研究

Study on Corneal thickness measurements with the Orbscan-Ⅱ Topography and ultrasonic pachymetry before and after LASIK
下载PDF
导出
摘要 目的研究Orbscan-Ⅱ角膜地形图仪对准分子激光原位角膜磨镶术(LASIK)手术前、后角膜厚度测量效果。方法应用Obscan-Ⅱ角膜地形图仪和A超角膜测厚仪分别测量150例(300只眼)LASIK手术前及术后6个月患者的角膜厚度,进行分析比较。结果术前A超角膜测厚仪测量结果平均为(545.43±29.079)μm,Orbscan-II测量结果平均为(535.17±32.352)μm,Orbscan-II比A超角膜测厚仪测量的中央角膜厚度薄(10.27±15.188)μm;术后6个月A超角膜测厚仪测量结果平均为(457.64±32.30)μm,Orbscan-II测量结果平均为(417.06±46.422)μm,Orbscan-II比A超角膜测厚仪测量的中央角膜厚度薄(51.15±18.4)μm。两参数检验有统计学意义,两种方法有高度相关性。结论Orbscan-II测量所得的角膜厚度值可以作为LASIK手术前后很好的参考依据。 Objective To investigate the difference on corneal thickness measurements using Orbscan-ⅡSystem and ultrasonic pachymetry before and after LASIK in myopic patients.Methods Orbscan-ⅡSystem and A-scan ultrasonography were used to measure the corneal thickness pre-and post-operative in 300 eyes of 150 cases that underwent myopic LASIK.The differences were evaluated by t tests.Results Pre-operatively,the mean corneal thickness was 535.17±32.352 μm with Orbscan-Ⅱsystem,and 545.43±29.079 μm with A-scan ultrasonography.Post-operatively,the mean corneal thickness with Orbscan-Ⅱsystem and with A-scan ultrasonography was 417.06±46.422 μm and 457.64±32.30 μm respectively,and the differences were statistically significant.Conclusion There is significant difference between the corneal thickness value measured by Orbscan-II topography system and ultrasonic pachymetry.Even high correlation between them has been showed,Orbscan-Ⅱsystem for corneal thickness measurement could not substitute A-scan ultrasonography completely.Certainly it still can be used as the very good reference for LASIK.
出处 《宁夏医学杂志》 CAS 2009年第12期1145-1146,共2页 Ningxia Medical Journal
关键词 准分子激光原位角膜磨镶术 Orbscan-Ⅱ A超角膜测厚仪 角膜厚度 laser in situ keratomileusis(LASIK) Orbscan II A-scan ultrasonography corneal thickness
  • 相关文献

参考文献4

二级参考文献19

  • 1Yaylali V,Kaufman SC,Thompsom HW. Corneal thickness measurements with the Orbscan Topography System and Ultrasound Pachymetry[J]. J Cataract Refract Surg,1997,23(9):1345-1350.
  • 2李凤鸣.眼科全书(上册)[M].北京:人民卫生出版社,1997.42.
  • 3Iskander NG,Anderson Penno E,Peter NT,et al. Accuracy of Orbscan pchymetry measurements and DGH ultrasound pachymetry in primary laser in situ Keratomileusis and LASIK enhancement procedures[J]. J Cataract Surg,2001,27(5):681-685.
  • 4Nissen J,Hjortdal JO,Ehlers N,et al. A clinical comparison of optical and ultrasonic pachymetry[J]. Acta Ophthalmol Copenh,1991,69(5):659-663.
  • 5Lattimore MR Jr. Influence of extended soft contact lens wear on the comparative measurenent of central corneal thickness[J]. Acta Ophthalmol Scand,1996,74(2):239-242.
  • 6Mertz GW. Overnight swelling of the living human cornea[J]. J Am Optom Assoc,1980,51(2):211-214.
  • 7Lattimore MR Jr,Sandor Kaupp MS,Schallhorn S,et al. Orbscan pachymetry:implication of a reperted measures and diurnal variation analysis[J]. Ophthalmology,1999,106(5):977-981.
  • 8Dursun D,Monroy D,Knighton R,et al. The effects of experimental tear film removed on corneal surface regularity and barrier function[J]. Ophthalmology,2002,107(9):1754-1760.
  • 9Seiler T,Koufala K,Richter G.Latrogenic keratectasia after laser in situ keratomileusis.Refract Surg,1998,14:312-317
  • 10Mc Corty CA,Aldred GF,Taylor HR.Comparison of results of excimer laser correction of all degrees of myopia at 12months postoperatively.Am J Ophthalmol,1996,121:372-380

共引文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部