Diaton眼压仪的临床应用 被引量:2

Clinical application of Diaton tonometer
下载PDF
导出
摘要 目的检测Diaton眼压仪的临床性能。方法分别使用Diaton眼压仪及Gold-mann压平眼压计测量103例(206眼)患者的眼压并分析结果的相关性。结果Diaton眼压仪和Goldmann压平眼压计测得的平均眼压分别为(16.46±2.03)mmHg(1kPa=7.5mmHg)和(18.04±2.63)mmHg,二者平均眼压比较,差异无统计学意义(F=0.091,P=0.918)。2种眼压仪测量值之差的平均值为(-1.05±2.12)mmHg,Diaton眼压仪测量平均值略低于Goldmann眼压计测量平均值,二者测量值的相关系数r=0.909,P=0.000,呈高度正相关。使用Goldmann压平眼压计的患者出现眼部异物感、畏光、流泪等不适,136眼出现荧光素染色阳性。使用Diaton眼压仪患者无不适,26眼角膜荧光素染色阳性。结论Diaton眼压仪测量值与Goldmann压平眼压计测量值高度正相关,且舒适度高,可在临床实践中推广使用。 Objective To investigate the clinical performances of Diaton tonometer(DT).Methods Intraocular pressure(IOP)was measured by DT and Goldmann applanation tonometer(GAT)on 206 eyes of 103 patients,and the correlation of IOP measured by DT and GAT were analyzed.Results The mean IOP by DT and GAT were(16.46±2.03)mmHg(1 kPa=7.5 mmHg)and(18.04±2.63)mmHg,respectively.There was no statistical difference in IOP between that by DT and that by GAT(F=0.091,P=0.918).The mean difference between these two kinds were(-1.05±2.12)mmHg, IOP by DT was lower than that by GAT, and there was a correlation between them(r = 0. 909,P = 0. 000). There were patients with ocular foreign body sensation, photophobia and lacrimation, and 136 eyes with positive fluorescein staining by GAT;No patient felt uncomfortable and 26 eyes were with positive corneal fluorescein staining by DT. Conclusion IOP measured by DT are highly correlated with that by GAT. DT has a high degree of comfort and can be used in clinical diagnosis.
出处 《眼科新进展》 CAS 北大核心 2009年第12期922-923,共2页 Recent Advances in Ophthalmology
关键词 青光眼 眼压测量 Diaton眼压仪 GOLDMANN压平眼压计 glaucoma intraocular pressure measurement Diaton tonometer Goldmann applanation tonometer
  • 相关文献

参考文献2

  • 1Heiman DC, Hodge DO, Boume WM. Change in corneal thickness in patient with treated and untreated ocular hypertension[J]. Cornea 2006;25(9) :639-643.
  • 2Toler MI, Vural A, Erdogan H, Topalkara A, Arici MK. Central corneal thickness and Diaton transpalpebral tonometry[J]. Graefes Arch Clin Exp Ophthalmol 2008 ;246(6 ) :881 -889.

同被引文献34

  • 1乔荣华,刘丽娜,李美玉,任泽钦.Tono-Pen眼压计临床应用评价[J].中国实用眼科杂志,2004,22(12):980-982. 被引量:12
  • 2张金顺,方爱武,彭寿雄,周业辉,潘伟华,王玉宏,王勤美.动态轮廓眼压计与Goldmann压平眼压计的临床应用比较[J].中国实用眼科杂志,2007,25(2):163-164. 被引量:2
  • 3Mohan S, Tiwari S, Jain A, et al. Clinical comparison of Pulsair? non- contact? tonometer? and Goldmann applanation? tonometer? in Indian population [J]. J Optom, 2014, 7 (2): 86-90.
  • 4Chai Y, Shao Y, Lin S, et al. Vision - related quality of life and emotional impact in children with strabismus: a prospective study [J]. J Int Med Res, "2009, 37 (4) : 1108 -1114.
  • 5Yiimaz I, Altan C, Aygit ED, et al. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non -contact airpuff tonometer, and Tono - Pen XL [ J ]. Clin Ophthalmol, 2014 (8): 1069-1074.
  • 6Ventura AC, Bfihnke M, Mojon DS. Central corneal thickness measurements in patients with normal tension glaucoma, primary open angle glaucoma, pseudoexfoliation glaucoma, or ocular hypertension [J]. Br J Ophthalmol, 2001, 85 (7): 792-795.
  • 7Shields MB. The non - contact tonometer. Its value and limitations [J]. Surv Ophthalrnol, 1980, 24 (4): 211-219.
  • 8Doherty MD, Carrim ZI, O'Neill DP. Diaton tonometry: an assessment of validity and preference against Goldmann tonometry [ J]. Clin Experiment Ophthalmol, 2012, 40 (4) : e171 -175.
  • 9Toker MI, Vural A, Erdogan H, et al. Central corneal thickness and Diaton transpalpebral tonoraetry[J]. Graefes Arch Chn Exp Ophthalmol, 2008, 246 (6): 881-889.
  • 10MOHAN S, TIWARI S, JAIN A, et al. Clinical comparison of pulsair non-contact tonometer and Goldmann applanation tonome- ter in Indian population[J]. J Optom, 2014, 7(2): 86-90.

引证文献2

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部