摘要 本文比较分析了三种x射线源(^(55)Fe、^(238)Pu和x光管)对木材进行微密度测定的精度,并对不同树种木材的质量衰减系数进行了测定和研究。 BY using direct scanning method, the accuracy of wood microdensitometry with 3 kinds of x-ray source ((55)~Fe, (238)~Pu and x-ray tube) was studied and the Variation of mass attenuation coefficients of 4 species of wood was determined and analized The measurement with (55)~Fe and x-ray tube was found more accurate than that with (238)~Pu because the x-ray energy of the formers is closer to the optimal one than the latter A certain degree of variation of wood mass attenuation coefficienat among species and within a species was showed in the tests, which should be considered as high accuracy of the measurement is expected
作者 梁景森 管宁
出处 《木材工业》 北大核心 1990年第4期33-37,共5页 China Wood Industry
关键词 木材 微密度 测定 X射线 X-ray microdensitometry Wood density Wood density variation Microdensitometrical characteristics Mass attenuation coefficient
  • 相关文献

参考文献1

  • 1D. J. Cown,B. C. Clement. A wood densitometer using direct scanning with X-rays[J] 1983,Wood Science and Technology(2):91~99