从质量控制的角度评价ASSET在腹部MR扫描中的应用价值 被引量:3

The Application Value of ASSET in Abdomen MR Scanning from A Quality-controlling Point of View
原文传递
导出
摘要 目的从质量控制的角度阐述阵列空间敏感性编码技术(array spatial sensitivity encoding technique,AS-SET)在腹部MR扫描中的应用价值。资料与方法20例患者分别进行采用ASSET和不采用ASSET的腹部常规扫描。选择肝右后叶和脾作为感兴趣区(ROI),测量信号强度,计算信噪比(SNR)和对比噪声比(CNR);伪影通过观察图像获得;分辨率指标由两名医师评估。结果CNR中轴位T1WI和冠状位快速平衡稳态进动序列(FIES-TA)两个序列使用ASSET和不使用ASSET时差异无统计学意义(P>0.05);SNR中两个序列使用ASSET和不使用ASSET时差异无统计学意义(P>0.05);伪影指标中两个扫描序列使用ASSET和不使用ASSET时差异有明显的统计学意义(P<0.05);分辨率指标中两个扫描序列使用ASSET和不使用ASSET时差异有显著统计学意义(P<0.05)。结论采用ASSET的影像伪影和分辨率对图像质量产生较大的影响,应尽可能少用或不用。 Objective To assess the application value of ASSET in abdomen MR images from a quality-controlling point of view. Materials and Methods Conventional abdomen MR scanning with and without ASSET were aquired in 20 patients. Regions of interest (ROI) were drawn on the right posterior lobe of liver and spleen. The signal intensity was measured. Signal noise ratio (SNR) and contrast noise ratio (CNR) were calculated. Artifact of the image was evaluated,Imag resolution was assessed by two radiologists. Results On axial T1 -weighted imaging and coronal FIESTA sequence, there was no statistical difference in CNR and SNR between sequences with ASSET and sequences without ASSET ( P 〉 0.05 ), but there was statistical difference in the artifact and imag resolution between sequences with ASSET and sequences without ASSET (P 〈 0.05 ). Conclusion Artifacts and image resolution of sequences with ASSET might have impact on images, which suggested ASSET was avoided to be used frequently.
出处 《临床放射学杂志》 CSCD 北大核心 2009年第12期1688-1691,共4页 Journal of Clinical Radiology
关键词 磁共振成像 质量控制 并行采集 腹部 Magnetic resonance imaging Quality-controlling Paralleling collecting Abdomen
  • 相关文献

参考文献8

  • 1Feinberg DS Rofsky NM,Johnson G. Multiple breath-hold averaging (tuba) method for increased SNR in abdominal MRI. Magn Reson Med, 1995,34:905.
  • 2Weiger M, Pruessmann KP, Leussler C, et al. Specific Coil Design for SENSE:A Six-Element Cardiac Array. Magn Reson Med,2001,45: 495.
  • 3Kyriakos WE, Panych LP, Kacher DF,et al. Sensitivity Profilesfrom an Array of Coils for Encoding and Reconstruction in Parallel ( SPACE RIP). Magn Reson Med.2000,44:301.
  • 4Weiger M, Pmessmann KP, Kassner A, et al. Contrast-enhanced 3D MRA using SENSE. JMRI,2000,12:671.
  • 5Zhu Y. Parallel excitalion with an array of transmit coils.Magn Reson Med,2004,51:775.
  • 6田俊,杨亚芳,韩晖云.MR成像伪影与消除[J].实用放射学杂志,2007,23(5):705-707. 被引量:14
  • 7唐磊,张晓鹏,孙应实,李英.胃癌磁共振扩散加权成像:并行采集联合分次屏气多信号平均技术的临床研究[J].中国医学影像技术,2005,21(12):1830-1834. 被引量:15
  • 8Pruessmann KP, Weiger M, Scheidegger MB, et al. SENSE : sensitivity encoding for fast MRI. Magn Reson Med, 1999 ,42:952.

二级参考文献7

  • 1郑晓林,徐辉雄,吕明德,许达生,徐作峰.磁共振弥散加权成像检测肝内小病灶的临床应用[J].中国医学影像技术,2004,20(10):1510-1512. 被引量:41
  • 2王世杰,罗立民.功能MRI的伪影校正方法[J].中国医疗器械杂志,2005,29(6):429-434. 被引量:1
  • 3燕树林.全国CT、MR技师上岗培训考试参考资料[M].北京:中华医学会继续教育部,1998.184-188.
  • 4Hashemi RH,Bradley WG,Lisanti CJ 著,尹建忠 译.MRI基础[M].第2版.天津:天津科技翻译出版公司,2004.209-235.
  • 5Aleeci M,Jezzard P.Characterization and reduction of gradient-induced eddy currents in the RF shield of a TEM resonator[J].Magn Reson Med,2002,48(2):404-407.
  • 6Papadakis NG,Martin KM,Pickard JD,et al.Gradient preemphasis calibration in diffution-weighted echo-planar imaging[J].Magn Reson Med,2000,44(4):616-624.
  • 7李晓红 卢光明.颈部磁共振血管造影的伪影及消除方法.中国医学影像技术,1999,15:32-33.

共引文献27

同被引文献18

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部