期刊文献+

血液净化系统的高可靠性漏血监测方法研究 被引量:4

Research on high reliably blood-leak monitoring in blood purification systems
下载PDF
导出
摘要 为提高血液净化系统中漏血监测的可靠性,减少漏血检测易出现的误报、漏报现象,提出一种可靠的漏血检测方法。采用小波分析技术对检测的光电转换信号进行处理,去除气泡通过时对漏血检测的噪声影响,并针对管壁污染的现象,采取漂移跟踪算法进行报警阈值自适应调整,以减少检测的误报和漏报。实验结果表明采用小波分析技术和漂移跟踪算法用于血液净化系统的漏血监测取得了较好的效果。 To improve the reliability of monitoring blood-leak in blood purification systems (BPSs), a new detecting approach for blood-leak is presented to reduce misinformation and report missing. Wavelet de-noising is adopted to eliminate the interference introduced by air bubbles in light-electronic transforming signal, and an algorithm tracking the detecting voltage drift due to dirty pipeline of waste fluid and adjusting the alarm threshold is used to enhance the adaptability of monitoring blood-leak. Experiments show that this method can be effectively applied in BPSs.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2009年第12期2640-2646,共7页 Chinese Journal of Scientific Instrument
基金 国家电子信息产业发展基金(工信部运[2008]97号) 重庆市重大科技攻关计划(CSTC2006AA5004)资助项目
关键词 血液净化系统 漏血监测 小波去噪 漂移跟踪 BPS blood-leak monitoring wavelet de-noising tracking drift
  • 相关文献

参考文献13

二级参考文献30

  • 1张喜成,孙斐.HG—Ⅰ型涂层厚度监测系统的监测仪设计[J].南京理工大学学报,1994,18(6):76-80. 被引量:3
  • 2郑树彬,林建辉,林国斌.用于轨检信号提取的线性相位IIR移变滤波器[J].西南交通大学学报,2007,42(2):211-216. 被引量:3
  • 3MALLAT S. Multifrequency channel decomposition of images and wavelet models[J].IEEE Transactions on Acoust. Speech Signal Process, 1989, 37 (12) : 2091-2110.
  • 4VIDAKOVIC B, LOZOYA C. On time-dependent wavelet denoising [ J ]. IEEE Transactions Signal Processing, 1998,46(9) :2549-2551.
  • 5DONOHO D, JOHNSTONE L. Ideal spatial adaptation via wavelet shrinkage [J].Biometrika, 1994, 81: 425- 455.
  • 6DONOHO D, JOHNSTONE L, KERKYACHARIAN G, et al. Wavelet shrinkage : asymptopia [ J ]. Journal of royal statistics society series(B) , 1995,57:301-369.
  • 7PAN Q, ZHANG L, DAI G, et al. Two denoising methods by wavelet transform [ J ]. IEEE Transactions Signal Processing, 1999,47 (12) : 3401-3406.
  • 8JANSEN M, BULTHEEL A. Multiple wavelet threshold estimation by generalized cross validation for images with correlated noise [J]. IEEE Transactions on Image Processing, 1999,8(7) :947-953.
  • 9WEYRICH N, WARHOLA T. Wavelet shrinkage and generalized cross validation for image denoising [ J ]. IEEE Transactions on Image Processing, 1998,7 ( 1 ) : 82-90.
  • 10CONLON E M. Bayesian models for pooling mieroarray studies with multiple sources of replications [ J ]. BMC Bioinformatics, 2006,7 : 247-261.