用于低能量除颤的除颤器设计 被引量:5

Design of a defibrillator used for low energy defibrillation
下载PDF
导出
摘要 设计了用于低能量除颤的,可灵活调节放电脉冲宽度和准确测量实际放电能量的除颤器。该除颤器的放电波形为双相指数截尾波,通过人工设置,可释放1-3组双相指数截尾波 放电脉冲的宽度、各脉冲之间的时间间隔可人工灵活调节,最小步长为0.1 ms 可测量放电前后储能电容电压,再根据储能电容的容值准确地计算出实际放电能量。该除颤器在双相指数截尾波的基础上,通过调节放电脉冲的宽度、各脉冲之间的时间间隔及脉冲数量,优化放电波形,达到低能量除颤的目的。已在20余例低能量除颤的动物实验中应用,各项功能符合设计要求。 A new defibrillator with adjustable shock duration and measurable shock energy has been designed. This defibrillator is based on biphasic truncated decaying exponential waveform and can release 1-3 groups of biphasic truncated decaying exponential waveform according to manual setup. The pulse duration and the interval between the pulses of the shock can also be set freely by the operator, and the minimum step is 0.lms. The voltages on the energy storage capacitor before and after the shock can be measured, and the accurate practical defibrillation energy also can be calculated from the voltages and the capacitance of the storage capacitor. This defibrillator optimizes the shock waveform to realize low energy defibrillation through adjusting the pulse duration, interval and number of the pulses on the basis of biphasic truncated decaying exponential waveform. The capabilities of the defibrillator are validated through more than 20 animal trial applications.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2009年第12期2664-2668,共5页 Chinese Journal of Scientific Instrument
基金 国家科技支撑计划(2007BA107A27) 上海市重点学科建设项目(B112)资助
关键词 体外除颤器 低能量除颤 放电脉冲宽度 实际放电能量测量 external defibrillator low energy defibrillation duration of defibrillation pulse measurement of practical defibrillation energy
  • 相关文献

参考文献11

  • 1宋二梅,邬小玫,杨翠微,方祖祥.双相指数截尾波体外除颤器的研制[J].中国医疗器械杂志,2006,30(1):25-28. 被引量:4
  • 2宋海浪,邬小玫,杨翠微,方祖祥.基于嵌入式系统的体外除颤器[J].仪器仪表学报,2006,27(z1):191-193. 被引量:3
  • 3Part 5: Electrical Therapies automated external defibrillators, defibrillation, cardioversion and pacing[J]. Circulation, 2005,112:Ⅳ-35-Ⅳ-46.
  • 4AMBLER J J S, SADO D M, ZIDEMAN D A, et al. The incidence and severity of cutaneous burns following external DC cardioversion[J]. Resuscitation 61 (2004):281-28.
  • 5BORIANI G, BIFFI M. Mechanisms of pain associated with intemal defibrillation shocks: Results of a randomized study of shock waveform[J]. Heart Rhythm 2005;2:708-713.
  • 6高蕴雅,龚镇楠,朱伯良,张觉先,孙友法,方祖祥,石镭,戚文航,王志刚.低能量心脏除颤心肌损伤的实验研究[J].中国心脏起搏与心电生理杂志,2001,15(2):113-115. 被引量:7
  • 7赖大坤 方祖祥.集成求解环境下胸腔电场分布的数值仿真分析.仪器仪表学报,2007,28(8):80-84.
  • 8LAID K, CAO T T, WU X M, et al. Optimization of electlodes displacement for transthoracic defibrillation: a simulation study on spatial configuration pattern[J]. International Journal of Bioelectromagnetism, 2007,9( 1):36-37.
  • 9LOWN B, NEUMAN J. Comparison of alternating current with direct current electroshock across the closed chest[J]. The American Journal of Cardiology, 1962:223- 233.
  • 10TACKER W A, GEDDES L A. The laws of electrical stimulation of cardiac tissue[J]. Proceedings of the IEEE, 1996:84(3):355-365.

二级参考文献22

  • 1宋二梅,邬小玫,杨翠微,方祖祥.双相指数截尾波体外除颤器的研制[J].中国医疗器械杂志,2006,30(1):25-28. 被引量:4
  • 2AHA,Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care:an international consensus on science.Circulation 2000,102(Suppl 1).
  • 3Schuder JC,et al.Thansthoracic ventricular defibrillation in the dog with truncated and untracted exponential stimuli.IEEE Trans BioMedEng 1971,18:410-415.
  • 4Schuder JC,et al.Transthoracic ventricular defibrillation with triangular and trapezoidal waveforms.Circ Res 1966;19:689-694.
  • 5Bossaert L,editor.European Resuscitation Council Guidelines for Resuscitation.Elsevier Science,1998.
  • 6Alrich U,et al.Waveforms of external defibrillators:analysis and energy contribution.Elsevier Science,1999.
  • 7Graig B.Clark,et al.Transthoeacic biphasic waveform defibrillation at very high and very low energy:a comparison with monophasic waveforms in an animal model of ventricular fibrillation;Resuscitation 54(2002)183-186.
  • 8Bardy GH,et al.Multicenter comparisons of truncated biphasic shocks and standard damped sine wave monophasic socks for transthoracic ventricular defibrillation,Circulation 1994,10:2507-14.
  • 9Poole JL, et al. Low energy impedance-compensating biphasic waveforms terminate ventricular fibrillation at high rate in victims of out-of-hospital cardiac arrest.J Clin Electrophysiol 1997,3:1373-1385.
  • 10Reddy RK,et al.Biphasic transthoracic defibrillation causes fewer ECG ST-segment changes after shock.Ann Emerg Med 1997,30:127-134.

共引文献8

同被引文献83

引证文献5

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部