埋伏牙低剂量多层螺旋CT检查辐射剂量的选择 被引量:1

原文传递
导出
摘要 目的探讨多层螺旋计算机体层成像(MSCT)在埋伏牙检查中的最低X射线剂量。方法回顾性分析我院60例埋伏牙十六排螺旋CT检查图像,将患者按CT管球每旋转一周X射线剂量不同划分为125mAs、25mAs、5mAs三组,由放射科两名正主任医师及三名副主任医师采用双盲法分析图像质量,并记录不同X射线剂量扫描时的CT权重剂量指数(CTDIw),进行统计学处理。结果①三组不同剂量扫描图像均能准确区分牙齿结构,5mAs扫描组在牙槽突骨小梁及周围软组织的分辨能力较25mAs、125mA组下降,但仍能达到诊断要求。②5mAs组CTDIw值仅为125mAs组的3.9%。结论5mAs的MSCT低剂量扫描可以满足埋伏牙检查需要,最大限度减少患者接受的辐射剂量。
出处 《中国辐射卫生》 2009年第4期491-492,共2页 Chinese Journal of Radiological Health
  • 相关文献

参考文献4

二级参考文献24

  • 1Freisfeld M,Dahl IA,Jager A,et al.X-ray diagnosis of impacted upper canines in panoramic radiographs and computed tomographs.J Orofac Orthop,1999,60: 177-184.
  • 2Checchi L,Alessandri Bonetti G,Pelliccioni GA.Removing high-risk impacted mandibular third molars:a surgical-orthodontic approach.J Am Dent Assoc,1996,127: 1214-1217.
  • 3Carol M, Petrina P,Graham JR.The radiographic localization of impacted maxillary canines: a comparison of methods[J].European Journal of Orthodontics,2001,23(1):25-34.
  • 4Freisfeld III, Dahl IA, Jagar A, et al. X-ray diagnosis of impacted upper canines in panoramic radiographs and computed tomograph[J].J Orthop,1999,60(3):177-184.
  • 5Bouquet A, Coudert JL, Bourgeois D, et al. Contributions of reformatted computed tomography and panoramic radiography in the localization of third molars relative to the maxillary sinus[J].Oral Surg Oral Med Oral Pathol Oral Radiol Endod,2004,98(3):342-347.
  • 6Smailiene D. Localization of impacted maxillary canines by palpation and orthopantomography[J].Medicina (Kaunas),2002,38(8):825-829.
  • 7Mason C, Papadakou P, Roberts GJ. The radiographic localization of impacted maxillary canines: a comparison of methods[J].Eur J Orthod,2001,23(1):25-34.
  • 8Jacobs SG. Radiographic localization of unerupted teeth: further findings about the vertical tube shift method and other localization techniques[J].Am J Orthod Dentofacial Orthop,2000,118(4):439-447.
  • 9Preda L, La Fianza A, Di Maggio EM, et al. The use of spiral computed tomography in the localization of impacted maxillary canines[J].Dentomaxillofac Radiol,1997,26(4):236-241.
  • 10Driessens FCM, Daculsi G. Tooth development and caries [J]. C RC Press, 1986, 1:4-20.

共引文献39

同被引文献10

  • 1王新怡,李福生.低剂量CT扫描在副鼻窦检查中的应用[J].中国辐射卫生,2005,14(4):315-317. 被引量:10
  • 2郭洪涛,刘勇,袁淑华.CT剂量指数(CTDI)测量研究[J].中国测试技术,2007,33(4):33-36. 被引量:33
  • 3Carol M, Petrina P, Graham JR. The radiographic localization o: impacted maxillary canines: a comparison of methods[J]. Eur J Orthodontics,2001,23(1) :25 34.
  • 4Adrian Becket A, Stella Claaushu S. Long-term follow up of se- verely resorhed maxillary incisors after resolution of an etiologi- cally associated impacted canine[J]. Am J Ortiaod Dentofaeial Or- thop,2005,127(6) :650 654.
  • 5Ballanti F, Lione R,Fiaschetti V, et al. Low dose CT protocol for orthodontic diagnosis[J]. Eur J Paediatr Dent, 2008,9 (2) : 65-70.
  • 6Kalra MK,Rizzo S,Maher MM,et al. Chest CT performed with Z- axis modulation: scanning protocol and radiation dose[J]. Radiolo gy, 2005,237 (1) : 303-308.
  • 7Slovis TL. CT and computed radiography: the pictures are great, but is the radiation dose greater than required[J]. Am J Roent- genol,2002,179(1) :39-41.
  • 8Stovis TL. The ALARA concept in pediatric CT:myth or reality.Radiology, 2002,4 (1) : 5-6.
  • 9Poiacin A, Kalender WA, Mareha| G. Evaluation o[ section sensi tivity profiles and image noise in spiral CT[J]. Radiology, 1992 185(1) :29-35.
  • 10贾明轩,张煊,刘为,王强.CT检查所致病人剂量的简便估算方法[J].中华放射医学与防护杂志,2001,21(5):374-376. 被引量:30

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部