摘要
目的:研究髋臼后柱拉力螺钉内固定技术中螺钉的最件进钉点、方向和长度。方法:取成年刃性髋骨标本20个,制作髋臼后住系列断面。测量进钉点与骨盆弓状线(OP)之距离,骶髂关节前端至P点间之距离(PQ),螺钉进入的长度,螺钉打入时的内倾角和前倾角;将测量数据输入到SPSS10.0进行统计学分析。结果:OP和PQ的距离分别为(16.8±2.1)mm和(23.5±3.4)mm,螺钉长度为(94.8±4.2)mm,螺钉内倾角17°18′±2°32′,前倾角57°36′±4°28′。结论:髋臼后柱拉力螺钉内匮I定技术具有创伤小、固定强度大等优点,在某些髋臼骨折中膀用具有明显的优势。
Objective:To investigate the best entry point, direction and length of screw in acetabular posterior column lag screw technique. Methods: Twenty male hip bone speeimens in adults were adopted to make serial cross - sections of acetabular posterior colunm. To detect the distances between entu point to linea areuata of pelvis minor (OP), P point and anterior extremity nf sacroiliac joint (PQ), and then to measure length of the inserted lag screw; the introverted and anteverted angle when a screw was inserted. These measured data were put into the SPSS 10.0 for statistical analysis. Results:The mean lengths of OP, PQ and inserted lag screw are (16.8±2.1) mm, (23.5±3.4) mm and (94.8±4.2) mm, respectively. The mean introverted and anteverted angle of screws are 17°18′±2°32′ and 57°36′±4°28′. Conclusions:The acetabular posterior column lag screw technique has advantages of little damage, high - intensity fixation, and so on. It has notable superiority in the treatment of |he fractures of aeetabulum.
出处
《解剖与临床》
2009年第6期417-419,共3页
Anatomy and Clinics
关键词
髋臼骨折
内固定
临床解剖
后柱螺钉
骨盆
Fracture of acetabulum
Internal fixation
Clinical anatomy
Posterior column lag screw
Pelvis