枕颈部三维运动范围的测量 被引量:7

A measurement on three dimensional movement of the occipitoatlantoaxial joint complex
下载PDF
导出
摘要 目的:研究枕颈部正常的三维运动范围,为枕颈部不稳分析和治疗提供生物力学基础。方法:运用脊柱三维运动测量分析系统,对11具成人新鲜枕颈部标本的三维运动范围进行测试。结果:寰枕关节的前屈、后伸、左/右侧屈、左/右轴向旋转运动范围分别是:11.7°、9.6°、2.8°/2.7°、6.9°/5.4°;寰枢关节分别为:8.9°、5.9°、3.4°/4.2°、37.5°/38.7°;寰枢关节中性区为27.4°,占主运动范围的71.7%;枕颈部与脊柱其它部位相比,主运动时伴有更明显的藕合运动,屈伸、侧屈时伴有比主运动范围更大的藕合旋转运动,分别为10.5°、14.3°。结论:脊柱三维测试分析系统不仅能对枕颈部的三维运动范围进行精确的测量,且能反映枕颈部三维运动的特点。 A measurement on the normal ranges of movement of the occipitoatlantoaxial joint complex was carried out on 11 fresh cadaver specimens of such joint complex to provide biomechanic base for the evaluating and treating its instability.A three dimensional spinal motion measurement and analysis system was used in this study.The motion range of flexion,extension,left/right lateral bending and left/right axial rotation were 11.7,9.6,2.8/2.7 and 6.9/5.4 degrees for atlantooccipital joint,8.9,5.9,3.4/4.2,and 37.5/38 degrees for atlantoaxial join respectively.The greatest intervertebral motion in the spine was axial rotation at the atlantoaxial joint,with the neutral zone constituting 71.7% of this motion.Meanwhile there was a significant coupling motion at atlantoaxial join.In flexion/extension,left/right lateral bending,the coupling rotation were 10.5, 14.3 degrees,and greater than the range of main motion.This study showed that the measurement was accurate and the specifity of mechanical behavior in occipitoatlantoaxial joint complex was acquired.
出处 《中国临床解剖学杂志》 CSCD 北大核心 1998年第4期360-362,共3页 Chinese Journal of Clinical Anatomy
关键词 枕颈部 三维运动 生物力学 测量 occipitoatlantoaxial joint complex three dimension movement biomechanics
  • 相关文献

参考文献4

二级参考文献3

  • 1Chang H,J Orthop Res,1992年,10卷,3期,446页
  • 2周定标,中华外科杂志,1991年,29卷,12期,73页
  • 3团体著者,实用神经外科学,1978年

共引文献11

同被引文献61

引证文献7

二级引证文献53

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部