螺旋CT的滤过算法影响图像质量的研究 被引量:3

Study of Influence of Spiral CT Filtering Algorithm on Image Quality
下载PDF
导出
摘要 目的探讨螺旋CT滤过算法对图像质量的影响,如何充分有效地利用滤过算法技术和使用规律。方法分析不同的滤过算法对扫描水模的空间分辨力、噪声、CT值的影响,再用不同滤过算法对人体不同部位组织器官的扫描图像进行对比分析。结果滤过算法对低对比分辨力和噪声的影响较大,肺算法的低对比分辨力最高,边缘算法的噪声最大,软组织算法最小;对空间分辨力骨加算法和边缘算法最高,噪声也最大;对软组织使用低对比和低噪声的算法,会使组织边缘更加圆滑,增加软组织对比,对骨组织和肺组织使用高空间分辨力算法会增加组织边缘的锐利程度,图像显示更加清晰。滤过算法对组织的平均CT值影响较小,但对最大和最小值有影响。结论恰当选择滤过算法是保证CT图像质量的重要条件。 Objective To discuss the influence of spiral CT filter algorithm on image quality,and discuss the method of effectively using filter algorithm and its orderliness.Methods Analyzed the resolution, noise, CT value of water phantom according to different reconstruction filter algorithm.Analyzed and compared the images of different area in the body by using filter algorithm.Results Low contrast resolution and noise were highly affected by filter algorithm. Low contrast resolution of lung algorithm was the highest.Noise of edge algorithm is high and of soft algorithm was small. Edge and bone-plus algorithm had a high spatial resolution and noise.Filter reconstruction algorithm of using low contrast resolution obtained soft tissue smooth boundary, sharply image was obtained through high resolution algorithm in bone tissue and lung tissue. It did not affect mean CT value badly, except max and min. Conclusion It is important to choose filter algorithm for ensuring precision of CT image quality.
作者 王厚军
出处 《中国医疗设备》 2010年第1期15-18,共4页 China Medical Devices
关键词 螺旋CT 滤过算法 图像质量 空间分辨力 spiral CT filter algorithm image quality spatial resolution
  • 相关文献

参考文献7

二级参考文献43

  • 1母其文,陈联盟,贾培沂,陈大经,王仁荣.回顾性重建在CT图像处理中的价值[J].中华放射学杂志,1994,28(11):785-786. 被引量:4
  • 2白玫,彭明辰.X线CT成像质量评估中的调制传递函数的测试与分析[J].医疗设备信息,2006,21(5):11-13. 被引量:15
  • 3[1]Kirillov AA: On a problem of I.M.Gel'fand. Sov Math Dokl 1961,2:268~269.(English translation)
  • 4[2]Smith BD: Cone-beam convolution formula. Comput Bio Med 1983, 13:81-87
  • 5[3]Tuy HK: An inversion for cone-beam reconstruction. SIAM ApplMath 1983, 43:546~522
  • 6[4]Grangeat P: Mathematical framework of cone beam 3d reconstruction via the first derivative of the radon transform. In:Herman GT,Louis AK and Natterer F,Eds. Mathematcal Methods in Tomography. Springer Verlag, 1990
  • 7[5]Smith BD: Image reconstruction from cone-beam projection:necessary and sufficient condition and reconstruction methods.IEEE Trans Med Imag 1985,MI-4: 14~28
  • 8[6]Axelsson C and Danielsson PE: Three-dimensional reconstruction from cone-beam data in time. Phys Med Biol 1994,39:477~491
  • 9[7]Edholm PR and Herman GT: Linograms in image reconstruction from projections. IEEE Trans MedicalImaging 1987, 6:301~307
  • 10[8]H.Kudo, F.Noo, and M.Defrise: Cone-beam filtered-backprojection algorithm fortruncated helical data. Phys. Med. Biol,1998, 43:2885~2909

共引文献48

同被引文献28

  • 1张博,龚建平.CT检查辐射致癌风险的研究进展[J].国际医学放射学杂志,2009,32(3):217-220. 被引量:39
  • 2张森,杜湘珂,李剑颖.心电图调制电流对于容积CT心脏扫描图像质量及曝光剂量影响的评估[J].中华放射学杂志,2006,40(9):974-976. 被引量:45
  • 3GB/T17589-1998,X射线计算机断层摄影装置影像质量保证检测规范[S].国家质量技术监督局,1999.
  • 4JJG1026-2007医用诊断螺旋计算机断层摄影装置(CT)X射线辐射源检定规程[S].北京:中国计量出版社,2007.
  • 5高建华,孙宪昶,李剑颖,李娜,夏庆堂,赵雯,戴汝平.不同前置滤线器对64层螺旋CT冠状动脉成像质量及放射剂量影响的对照研究[J].中华放射学杂志,2007,41(8):858-861. 被引量:51
  • 6JJG961-2001,医用诊断计算机断层摄影装置(CT)X射线辐射源检定规程[s].北京:中国计量出版社,2001.
  • 7医疗器械临床使用安全管理规范(试行)[s].北京:国家卫生部.2010.
  • 8Brian P. Shapiro,Phillip M. Young,Birgit Kantor,Yeon Hyeon Choe,Cynthia H. McCollough,Thomas C. Gerber.Radiation Dose Reduction in CT Coronary Angiography[J]. Current Cardiology Reports . 2010 (1)
  • 9Michael L. Steigner,Hansel J. Otero,Tianxi Cai,Dimitrios Mitsouras,Leelakrishna Nallamshetty,Amanda G. Whitmore,Hale Ersoy,Noah A. Levit,Marcelo F. Di Carli,Frank J. Rybicki.Narrowing the phase window width in prospectively ECG-gated single heart beat 320-detector row coronary CT angiography[J]. The International Journal of Cardiovascular Imaging . 2009 (1)
  • 10Sebastian Leschka,Paul Stolzmann,Florian T. Schmid,Hans Scheffel,Bjoern Stinn,Borut Marincek,Hatem Alkadhi,Simon Wildermuth.Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose[J]. European Radiology . 2008 (9)

引证文献3

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部