磁性附着体的衔铁对磁共振成像的影响 被引量:2

Impact of magnetic attachment keepers on magnetic resonance imaging examination
下载PDF
导出
摘要 目的:研究磁性附着体的衔铁对磁共振成像的影响。方法:体外实验选用MAGFIT EX400,MAGFITEX600及MAGNEDISC 800型磁性附着体的衔铁,分别设计为不铸接、与镍铬合金铸接和与金钯合金铸接3种形式,并完成衔铁根帽的制作。体内实验选择需行磁性附着体可摘义齿修复的患者作为研究对象。用SiemensSONATA1.5-T磁共振仪对所有样本及研究对象进行扫描,对体外样本伪影最大径进行测量和比较,对体内实验的伪影对头颈部解剖结构成像的影响进行分析。结果:MAGFIT EX400型、MAGFIT EX600型及MAGNEDISC 800型衔铁的伪影最大纵径分别为158.94、168.52和173.00 mm。衔铁伪影对颏部、舌、唇、软硬腭、颌骨的磁共振成像造成明显影响,位于后牙区的衔铁对脑及脊髓的成像造成影响。采用SE序列及左右向的频率编码,可以减小衔铁对头颈部MRI的影响。结论:磁性附着体的衔铁对磁共振成像有明显影响。 Objective:To evaluate the effects of magnetic attachment keepers on MRI images.Me-thods:In in vitro part,keepers of MAGFIT EX 400,MAGFIT EX 600 and MAGNEDISC 800 magnetic attachment systems were cast into standard root-caps with different alloy respectively,including nickel-chromium alloy and gold-palladium alloy,or not be cast for contrast.In in vivo part,volunteers with keepers in different position and amount were included.All the specimens and volunteers were imaged by a Siemens SONATA 1.5-T MRI scanner.Extent of the artifacts in every slice was measured.The magnitude of the artifacts and the distortion of the anatomic structures were compared.Results:All kinds of keepers being studied produced obvious MRI artifacts.The extent of MRI artifacts induced by MAGFIT EX 400 keepers,MAGFIT EX 600 keepers and MAGNEDISC 800 keepers are 158.94 mm,168.52 mm and 173.00 mm,respectively.The images of mental region,tongue,palate,jawbone and sinus were evidently obscured by artifacts in all cases.When keeper was put in the molar region,the keeper-related MRI artifact may obscure the imaging of brain,spinal cord.SE sequence with right-left frequency-encoding direction was more desirable in MRI for the patients with magnetic attachment keepers.Conclusion:Keepers of magnetic attachment system did induce obvious artifacts in MRI examination.
出处 《北京大学学报(医学版)》 CAS CSCD 北大核心 2010年第1期67-73,共7页 Journal of Peking University:Health Sciences
关键词 磁力学 磁共振成像 义齿 局部 可摘 Magnetic Magnetic resonance imaging Denture partial removable
  • 相关文献

参考文献4

  • 1Shellock FG, Crues JV. High-field strength MR imaging and metallic biomedical implants : an ex vivo evaluation of deflection forces [J]. MR, 1988, 151(2) : 389 -392.
  • 2刘玉华,孙樱琳.固定义齿修复材料对MRI图像的影响[J].现代口腔医学杂志,2005,19(6):588-589. 被引量:19
  • 3Ernstberger T, Heidrieh G, Buchhorn G, et al. Postimplantation MRI with cylindric and cubic intervertebral test implants: evaluation of implant shape, material, and volume in MRI artifacting- an in vitro study[J]. Spine J, 2007, 7(3) : 353 -359.
  • 4Florentina T. Magnetic resonance imaging artifact and the magnetic attachment system[J]. Dent Mater J, 1994, 13 (1) : 76-88.

二级参考文献9

  • 1New PF,Rosen BR,Brady TJ,et al.Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging.Radiology,1993,147:139-148.
  • 2Kanal E,Shellock FG,Talagala L.Safety considerations in MR imaging.Radiology,1990,176:593-606.
  • 3Shellock FG,Shellock VJ.Metallic stents:evaluation of MR imaging safety.AJR,1999,173:543-547.
  • 4李联忠 戴建平 赵斌.颅脑MRI诊断与鉴别诊断[Z].,..
  • 5Bui FM,Bott K,Mintchev MP.A quantitative study of the pixel-shifting,blurring and nonlinear distortions in MRI images caused by the presence of metal implants.J Med Eng Technol,2000,24:20-27.
  • 6Thomsen M,Schneider U,Breusch SJ,et al.Artefacts and ferromagnetism dependent on different metal alloys in magnetic resonance imaging.An experimental study.Orthopade,2001,30:540-544.
  • 7Takashi K,Manabu M,Hugh D,et al.Dental bur fragments causing metal artifacts on MR images.AFNR Am J Neuroradiol,1998,19:317-319,13-16.
  • 8Fache JS,Price C,Hawbolt EB,et al.MR imaging artifacts produced by dental materials.AJNR Am J Neuroradiol,1987,8:837-840.
  • 9Hinshaw DB Jr,Holshouser BA,Engstrom HIM,et al.Dental material artifacts on MR images.Radiology,(缺卷期页)

共引文献18

同被引文献36

  • 1刘玉华,孙樱琳.固定义齿修复材料对MRI图像的影响[J].现代口腔医学杂志,2005,19(6):588-589. 被引量:19
  • 2汪厚希,丁仲鹃,林云红,周屹立,谢静.种植磁附着体在全口义齿修复中的临床应用[J].华西口腔医学杂志,2005,23(6):515-517. 被引量:9
  • 3郭克熙,杜莉.磁性附着体衔铁耐腐蚀性能的研究进展[J].临床口腔医学杂志,2007,23(5):315-317. 被引量:3
  • 4徐世同,程祥荣,华先明,程祥惠.磁固位与杆固位种植全口义齿的三维有限元应力分析[J].中华口腔医学杂志,1997,32(2):105-107. 被引量:22
  • 5Gillings BR. Magnetic denture retention systems: inexpensive and efficient[J]. Int Dent J, 1984, 34 (3):184-197.
  • 6Tomoya G, Yoshinobu M. Why are magnetic attachments popular in Japan and other Asian countries[J]. Japan Dent Sci Rev, 2011 (47): 124-130.
  • 7Takahashi N, Takada Y, Okuno O. Galvanic corrosion between dental precious alloys and magnetic stainless steels used for dental magnetic attachments[J]. Dent Mater J, 2008, 27(2):237-242.
  • 8Sato K, Yamaguchi H, Miyamoto H, et al. Growth of human cultured cells exposed to a non-homogeneous static magnetic field generated by Sm-Co magnets [J]. Biochim Biophys Acta, 1992, 1136 (3):231-238.
  • 9Heckmann SM, Winter W, Meyer M, et al. Overdenture attachment selection and the loading ofimplant and denture-bearing area. Part 2: a methodical study using five types of attachment[J]. Clin Oral Implants Res, 2001, 12(6):640-647.
  • 10Takeshita S, Kanazawa M, Minakuchi S. Stress analysis of mandibular two-implant overdenture with different attachment systems[J]. Dent Mater J, 2011, 30(6): 928-934.

引证文献2

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部