The performance test is an important and necessary work for the micro-CT (computed tomography) system. The focal spot size of the micro focus X-ray tube is measured. The method of measuring the spatial resolution of micro-CT is introduced. A line-pair resolution of 28.2 lp/mm at the 10% modulation transfer function (MTF) level can be achieved with 14.7 μm spot size, 12.3 μm voxel size and a 25 mm field of view. In addition, a tungsten wire with the diameter of 5 μm can be detected by the system.
The performance test is an important and necessary work for the micro-CT (computed tomography) system. The focal spot size of the micro focus X-ray tube is measured. The method of measuring the spatial resolution of micro-CT is introduced. A line-pair resolution of 28.2 lp/mm at the 10% modulation transfer function (MTF) level can be achieved with 14.7 μm spot size, 12.3 μm voxel size and a 25 mm field of view. In addition, a tungsten wire with the diameter of 5 μm can be detected by the system.