期刊文献+

Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system 被引量:4

Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system
下载PDF
导出
摘要 In this paper,the beam quality and percent depth dose curves for different field sizes of CyberKnife? system were investigated by Monte Carlo simulations using the PENELOPE code,which has been used to simulate 6 MV photon beam.In water phantom,the absolute doses were calculated for Φ10–60 mm collimators,and percent depth dose curves were evaluated for Φ30–60 mm collimators.The agreement of dose distributions of the calculation with measurement was within 3.0%.The mean energy of photon spectrum was 1.46 MeV,and the beam quality index was 0.632,which was slightly smaller than that of measurement. In this paper, the beam quality and percent depth dose curves for different field sizes of CyberKnife system were investigated by Monte Carlo simulations using the PENELOPE code, which has been used to simulate 6 MV photon beam. In water phantom, the absolute doses were calculated for q,10~0 mm collimators, and percent depth dose curves were evaluated for q)30-60 mm collimators. The agreement of dose distributions of the calculation with measurement was within 3.0%. The mean energy of photon spectrum was 1.46 MeV, and the beam quality index was 0.632, which was slightly smaller than that of measurement.
出处 《Nuclear Science and Techniques》 SCIE CAS CSCD 2010年第1期16-19,共4页 核技术(英文)
基金 Supported by Shanghai Science and Technology Committee (08JC1410200) Shanghai Leading Academic Disciplines (S30109)
关键词 光子束 蒙地卡罗 外科治疗 立体定向 系统 PENELOPE 放射 光束质量 Monte Carlo, PENELOPE, CyberKnife, Beam quality, Percent depth dose
  • 相关文献

参考文献19

  • 1Adler J R, Chang S D, Murphy M J, et al. Stereotact Funct Neurosurg, 1997, 69: 124-128.
  • 2Shiomi H, Inoue T, Nakamura S, et al. Radiat Med, 2000, 18(2): 107-113.
  • 3Chang S D, Main W, Martin D P, et al. Neurosurg, 2003, 52:140-146.
  • 4Deng J, Guerrero T, Ma C M, et al. Phys Med Biol, 2004,49:1689-1704.
  • 5Salvat F, Fernandez-Varea J M, Sempau J. PENELOPE- 2006: A Code System for Monte Carlo Simulation of Electron and Photon Transport, Barcelona, 2006, 4-7.
  • 6Fujio A. Med Phys, 2006, 33(8): 2955-2963.
  • 7Yamamoto T, Teshima T, Miyajima S, et al. Radiat Medic, 2002, 20(6): 305-310.
  • 8Lymperopoulou G, Petrokokkinos L, Papagiannis P, et al. Nucl Instrum Meth Phys Res A, 2007, 580: 548-551.
  • 9Castellano G, Brusa D, Carrara M, et al. Nucl Instrum Meth Phys Res A, 2007, 580: 502-505.
  • 10MCNP - A General Monte Carlo N-particle transport code, http://mcnp-green.lanl.gov/.