植入型心律转复除颤器导致误放电8例

下载PDF
导出
摘要 目的探讨植入型心律转复除颤器(ICD)导致的误放电原因,为避免和减少误放电提供参考。方法分析8例患者共植入11台ICD的临床随访资料。8例患者7例为男性,1例女性,其中单腔ICD 7台,双腔ICD 3台,三腔植入型心律转复除颤器(CRT-D)1台。随访2~12年,平均7年。结果随访中除1例死亡外其余患者均存活。ICD共检出心律失常568次,启动治疗程序125次,其中误放电16次(12.8%),大部分发生在早年植入ICD的3例患者中。误放电原因包括:感知窦性心动过速3次,占误放电的18.8%,电磁干扰4次,占25%,电极导线绝缘层破损干扰9次,占56.3%。结论ICD误放电会给患者带来极大的身心伤害,应该受到重视。
出处 《广东医学》 CAS CSCD 北大核心 2010年第3期353-355,共3页 Guangdong Medical Journal
  • 相关文献

参考文献11

  • 1MOSS A J, HALL W J, CANNOM D S, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Inverstigators (MAD1T) [ J ]. N Engl J Med, 1996, 335(26) : 1933 - 1940.
  • 2Anonymous. A Comparison of antiarrhythmic - drug therapy with implantable defibrillators in patients resuscitated from near- fatal ventricular arrhythmias. The Antiarrhythmics versus Implanted Defibrillator(AVID) Investigators [ J ]. N Engl J Med, 1997, 337 (22) : 1576 - 1583.
  • 3王方正,华伟,张澍,陈新.69例植入型心律转复除颤器治疗患者的随访[J].中华心律失常学杂志,2001,5(2):120-122. 被引量:34
  • 4DORWARTH U, FREY B, DUGAS M, et al. Transvenous defibrillation leads : high incidence of failure during long - term folloe - up[ I]. J Cardiovasc Electrophysiol, 2003, 14( 1 ) : 38 -43.
  • 5刘平.影响埋藏式心脏起搏器寿命的“先天”因素[J].中国心脏起搏与心电生理杂志,2003,17(3):224-226. 被引量:1
  • 6WASHIZUKA T, CHINUSHI M, TAGAWA M, et al. Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias [ J ]. Jpn Circ J, 2001, 65(7) : 927 -930.
  • 7DURU F, BAUERFELD U, CANDINAS R. Avoiding inappropriate ventricular tachyeardia detection due to T - wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrograrn width criterion[J]. Europace, 2001, 3(1 ) : 80 -84.
  • 8GOLLOB M H, SEGER J j. Current status of the implantable cardioverter - defibrillator[ J]. Chest, 2001, 119(4) : 1210 - 1221.
  • 9胡雪松,方卫华,衣卫民,刘幼文.经静脉植入心律转复除颤器并四年随访(摘要)[J].中华心律失常学杂志,1998,2(2):98-98. 被引量:1
  • 10MANOLIS A G, KATSIVAS A G, VASSILOPOULOS C V, et al. Implantable cardioverter defibrillator - an unusual case of inappropriate discharge during showering[ J]. J Interv Card Electrophysiol, 2000, 4(1): 265 -268.

二级参考文献21

  • 1Oham OJ,Danilovic D. Improvements in pacemaker energy consumption and functional capability : four decades of progress [ J ]. PACE ,1997,20:2.
  • 2Kindermann M, Schwaab B, Berg M, et al. Longevity of dual chamber pacemaker: device and patient related determiants [ J ]. PACE, 2001,24(5) :810.
  • 3Glikson M, Hayes DL. Cardiac pacing: A review [ J]. Medical-Clin-North-Am,2001 ,85(2) :369.
  • 4Parsonnet V. Berstein ADPerry GY The nuclear pacemaker:is renewed interest warranted [ J] ? Am J Cardioal, 1990,66(10) :837.
  • 5Markewitz A, Kronski D, Kammeyer A, et al. Determinants of dual chamber pulse generators longevity [ J ]. PACE, 1995,18 ( 12 Pt 1 ) :2116.
  • 6Schuchert A, Meinertz T. A randomized study on the effects of pacemaker programming to a lower output on projected pulse generation longevity[ J]. PACE ,2001,24(8 pt 1 ):1 234.
  • 7Schwaab B, Schwerdt H, Heisel A, et al. Chronic ventricular pacing using an output amplitude of 1.0 volt[ J]. PACE , 1997, 20(9 pt 1 ) :2171.
  • 8Coates S,Thwaites B. The strenth-duration curve and its importance in pacing efficiency: a study of 325 pacing leads in 229 patients [ J].PACE ,2000, 23(8) :1 273.
  • 9Furman S, Garvey J, Hurzeler P. Pluse duration variation and electrode size as a factor in pacemaker longevity[ J]. J Throac Cardiovasc Surg,1975,69(3) :382.
  • 10Grossly GH, Brinker JA, Reynolds D, et al. Steroid elution improves the stimulation threshold in an active-fixation atrial-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 investigators[ J]. Circulation, 1995, 92(10) :2 935.

共引文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部