浅析F30-IIG200mA X线机主可控硅的单向触发

Analysis of Mono Flip-Flop in the Silicon-Controlled Rectifier of F30-IIG 200mA X-Machine
下载PDF
导出
摘要 近几年北京万东公司出厂的F30-IIG200mA X线机与以往出厂的该型机器在电路上有所改进,其中高压初级电路由原来接触器控制通断改为由可控硅控制的无触点开关去控制高压初级通断,当可控硅有一路损坏开路或未触发时,使高压初级电流猛增、自耦变压器发出嗡嗡过载负荷声、拍片时厚部位不能穿透,负载时千伏表、电压表数值下跌明显。 F30-ⅡG200mA X machine has been greatly improved in the electric circuits comparing with the former types of the machine produced by Beijing Wandong Company,especially the turning on and off high pressure junior electric circuits has been controlled by non-touching button monitored by silicon-controlled rectifier instead of touching-controller. As soon as the damage or non-touching occurs during the electric current circuits of silicon-controlled rectifier,the current of high pressure junior electric circuits inereases abruptly,the noise of over-loaded current lets out from photo-coupler capacitor,x-ray penetration cannot reach the thick part when photographing, the number of volt and voltage decrease significantly when electric current over-loaded.
作者 濮晓明
出处 《中国现代医生》 2010年第8期97-97,102,共2页 China Modern Doctor
关键词 脉动半波整流 磁通密度 磁场强度 峰值 有效值 Pulse half-way rectifier Magneto motive dense Magnetic field density Peak point Valid number
  • 相关文献

参考文献2

  • 1黎式棠.医用X线机原理与维修[M].南宁:广西科学技术出版社,1999.
  • 2赵文银.浅谈F78-300毫安X线机主可控硅单向触发.中外医用放射技术,1991,:30-30.