期刊文献+

心脏永久起搏器置入患者接受磁共振检查安全性探讨 被引量:1

Investigation on safety of magnetic resonance imaging on patients with permanent cardiac pacemakers
下载PDF
导出
摘要 目的探讨心脏永久起搏器非依赖患者接受磁共振成像(MRI)检查的安全性。方法选择7例起搏器非依赖患者分别行头、腰椎、髂血管和心脏MRI检查。检查前、后对生命体征进行记录,并作统计学分析,随访心电图。结果7例患者检查中无自述不适。6例患者未见心律失常及起搏器感知、起搏功能异常。1例患者检查中(射频脉冲启动时)出现自身缓慢心率,心率55次/分左右,偶见起搏QRS波形,但患者神志清楚,未诉心悸、不适,生命体征正常,在专科医生密切观察及严密心电监护下顺利完成检查。7例患者检查后无自主不适,随访心电图示起搏器功能正常。结论心脏永久起搏器非依赖患者可安全接受MRI检查。 Objective To investigate the safety of magnetic resonance imaging (MRI)on the patients with permanent cardiac pacemakers. Methods Seven patients with permanent cardiac pacemaker(non-dependent) underwent MRI scans of head,lower baek or heart. Continuous monitoring was performed during MRI scan including ECG and the vital signs. To follow up the vital signs be- fore,after scan and 1 week later. ECGs performed before and after MRI scan immediately. Statistical analysis was performed on da- ta. Results Seven patients ecomplained of uncomfort during or after MRI scan. Continuous monitoring was normal in 6 patients, 1 patient develuped slow hearl rale, FTR 55 heals per minute, occasional wide QRS,no palpitation,normal vital signs,finished MRI scan under the monitor of dcctors, All the patients had no complain after MR1, follow up ECGs and the pacemakers function were normal. Conclusion MRI can he safely performed on the patients with permanent cardiac pacemakers(non-dependent).
出处 《重庆医学》 CAS CSCD 北大核心 2010年第6期699-700,702,共3页 Chongqing medicine
关键词 心脏永久起搏器 非依赖者 磁共振 安全性 cardiac pacemaker non dependenl magnet ic resonance imaging safety
  • 相关文献

参考文献8

  • 1王方正,张澍,华伟,胡大一,陈新,蒋文平.全国心脏起搏器临床应用调查(2002至2005年)[J].中国心脏起搏与心电生理杂志,2007,21(3):272-275. 被引量:15
  • 2陈光辉,李航,付天果.永久性心脏起搏器植入的并发症及防治体会[J].重庆医学,2007,36(18):1870-1871. 被引量:5
  • 3Palicek W,Geisinger M, Castle I., et al. The effect of nuclear magnetic resonance on patients with cardiac pacemakers[J]. Radiology, 1983,147 ( 1 ) : 149.
  • 4Martin ET,Coman JA, Shellock FG, et al. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla [J]. J Am Coll Cardiol,2004,43(7) : 1315.
  • 5Barnes E, Woods T. High-field MRI of the brain safety acquired in pacemaker patients[EB/OL]. (2006-10-15). http ://Aunt Minnie. com.
  • 6Halperin H. MRIs made safe for people with modern defibrillators and pacemakers [EB/OL]. ( 2006-09-060918163607 ). http://www. sciencedaily. com/releases/2006/09/060918163607/html.
  • 7Luechinger R, Duru F, Zeijlemaker VA, et al. Pacemaker reed switch behaivior in 0.5,1.5 and 3.0 Tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields? [J]. PACE, 2002, 25 (10): 1419.
  • 8Gimbel JR, Johnson D, Levine PA, et al. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers[J]. PACE, 1996,19(6) :913.

二级参考文献4

共引文献18

同被引文献8

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;