脊柱椎弓根螺钉内固定术后行多层螺旋CT扫描时减少伪影及放射剂量的参数优化研究 被引量:8

Optimization of parameters for reducing artifact and radiation dose in multi-detector computer tomography after pedicle screw internal fixation
下载PDF
导出
摘要 目的评价多层螺旋CT检查时管电压和管电流与金属伪影程度之间的关系,探索适当的方案以减少金属伪影和辐射剂量。方法将4枚椎弓根螺钉植入人体腰椎L4、L5标本内制成的椎弓根螺钉内固定模型用丝线悬吊固定于伪影评价专用水模中央。对其使用不同扫描参数(主要是管电压及管电流)条件下的多层螺旋CT扫描,并测量椎弓根螺钉前端的放射状伪影长度以评价其伪影程度,记录辐射剂量。结果多层螺旋CT管电流的变化并没有明显影响金属伪影,但管电压的变化对伪影程度有较大影响,随着管电压的增大(100~140kV),伪影程度明显减小(P<0.05)。辐射剂量随着管电流及管电压的升高而升高。结论在其他扫描参数不变,能同时减少伪影和辐射剂量的方案是:高管电压(一般为140kV)并保持不变,而管电流则使用低于250~270mA常规使用值的电流。 Objective To assess the correlation between the metal artifact degrees with the tube voltage(in peak kilovolts,kV) and tube current(in milliamperes,mA) in multidetector computed tomography(MDCT),so as to search for proper strategies to reduce metal artifact and the radiation dose.Methods Four lumbar pedicle screws were implanted in the human L4,L5 specimens to establish the pedicle screw fixation model,and the models were hung on the appropriative water mode to assess the artifacts with silk thread.Different scanning parameters(Hnly tube tension,tube current) were used for scanning.The lengths of the artifacts emanating from the anterior tips of the screw were used to evaluate the degree of artifact.Results The tube current change did not significantly affect the metal artifacts(P0.05),but the tube voltage had significant impact on the metal artifacts;the degree of artifacts decreased remarkably with the increase of the tube voltage(100-140 kV,P0.05).Radiation dose rose with tube current and tube voltage increase.Conclusion Simultaneous reduction of the artifact and radiation dose can be achieved with high tube voltages(usually 140 kV) and tube currents should be lower than the generally used values of 250-270 mA.
出处 《第二军医大学学报》 CAS CSCD 北大核心 2010年第3期242-245,共4页 Academic Journal of Second Military Medical University
基金 国家自然科学基金(30872611) 上海市科委基础研究项目(07jc14069)~~
关键词 螺旋计算机体层摄影术 图像质量 伪影 辐射剂量 植入物 椎弓根螺钉内固定术 管电压 管电流 扫描参数 放射剂量 spiral computed tomography image quality artifacts radiation dosage implantation
  • 相关文献

参考文献18

  • 1Ohashi K, El Khoury G Y, Bennett D L, Restrepo J M, Berbaum K S. Orthopedic hardware complications diagnosed with multi detector row CT[J]. Radiology, 2005 ; 237 : 570-577.
  • 2Dalai T,Kalra M K,Rizzo S M,Schmidt B,Suess C,Flohr T,et al. Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients[J]. Radiology, 2005,236 : 671-675.
  • 3Kalra M K,Maher M M,Toth T L, Hamberg L M,Blake M A, Shepard J A, et al. Strategies for CT radiation dose optimization [J]. Radiology, 2004,230 : 619-628.
  • 4Hundt W,Rust F,Stabler A,Wolff H,Suess C,Reiser M. Dose reduction in multislice computed tomography[J]. J Comput Assist Tomogr,2005,29 :140-147.
  • 5王时庆,吴新淮,常钧,耿西亮.CT图像伪影的产生及消除[J].医疗卫生装备,2002,23(5):59-59. 被引量:10
  • 6时劲松.影响CT影像噪声的相关因素[J].中华放射医学与防护杂志,2004,20(2):112-113.
  • 7Prokop M. General principles of MDCT[J].Eur J Radiol,2003, (45 Suppl 1):S4-S10.
  • 8Sahani D,Saini S,D'Souza R V,O'Neill M J,Prasad S R,Kalra M K,et al. Comparison between low (3 ." 1) and high (6 = 1) pitch for routine abdominal/pelvic imaging with multislice computed tomography[J]. J Comput Assist Tomogr, 2003,27: 105-109.
  • 9Shrimpton P C, Hillier M C, Lewis M A,Dunn M. NRPB W67: Doses from computed tomography (CT)examinations in the UK-2003 review [M]. Oxon: National Radiologicai Protection Board, 2005 : 32.
  • 10Yazdi M,Gingras L,Beaulieu L. An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: experimental and clinical studies[J].Int J Radiat Oncol Biol Phys,2005,62:1224-1231.

二级参考文献11

  • 1[5]IEC 1223-2-6 constancy test-X-ray equipment for computed tomography. Interationl electrotechnical commission, 1994, 11
  • 2[6]Crespi A, Leoni S, Montanari G,et al. The evaluation of the physical characteristics of a volumetric computer tomography. Radiol Med (Torino), 1996, 91:460
  • 3[8]Verellen D, Vinh-Hung V, Bijdekerke P,et al. Characteristics and clinical application of a treatment simulator with Ct-option. Radiother Oncol, 1999, 50:355
  • 4[9]McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys, 1999, 26:2223
  • 5[1]Smith A, Shah GA, Kron T. Variation of patient dose in head CT. Br J Radiol, 1998, 71:1296
  • 6[2]Hidajat N, Schroder RJ, Wolf M,et al. Dosage values for characterization of patient exposure in computerized tomography and their significance for risk assessment. Radiology, 1997, 37:464
  • 7[3]Heinz-Peer G, Weninger F, Nowotny R,et al. Radiation dosage of various CT techniques in diagnostic lung imaging. Radiology, 1996, 36:470
  • 8Berrington de Gonzalez A,Darby SS. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet, 2004,363(9406):345-351.
  • 9United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation. New York:United Nations, 2000.
  • 10尉可道.CT剂量的测量及其表达[J].中华放射医学与防护杂志,1997,17(5):347-351. 被引量:43

共引文献24

同被引文献68

引证文献8

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部