RayPex5根尖定位仪测量根管工作长度准确性的临床评价 被引量:5

A clinical evaluation on accuracy of RayPex5 apex locator for measuring root canal working length
下载PDF
导出
摘要 采用手感法、电测法测量211个根管工作长度,而后用X线片法对所确定的根管工作长度进行验证。电测法和手感法准确率分别为96.21%和78.20%,两者有显著统计学差异(P<0.001)。电测法测量前牙和磨牙的准确率分别为100%和92.5%,两者有统计学差异(P<0.05),而测量根尖周有病变与无病变组、前牙与前磨牙组、前磨牙与磨牙组准确率均无统计学差异。研究表明,电测法测定根管工作长度准确性高,且测量前牙的准确率高于测量磨牙的准确率;Raypex5型根尖定位仪可以较准确地测量根管工作长度。 In order to evaluate the accuracy of RayPex5 apex loeator in measuring root canal working length in elinical practice and to analyze the possible factor resulting in deviation, the working length of total 211 root canals were measured by handing method and electronic method respectively. The exact working length of those root canals were determined by X-ray method. The accuracy rate of electronic method and handing method was 96.21% and 78.2% respectively, with significant difference(P 〈0. 001 ). The accuracy rate of front teeth group and molar group measured by electronic method was 100% and 92.5% respectively, with significant difference (P 〈 0.05 ) , while the apical paradentitis group and the apical non-paradentitis group, the front teeth group and the premolar group, and the premolar group and the molar group didn't differ from each other, respectively. The study showed that the accuracy rate of electronic method was high for determining root canal working length, and Raypex 5 apex loeator could define apical seat to measure root canal working length accurately.
出处 《实用口腔医学杂志》 CAS CSCD 北大核心 2010年第2期260-261,共2页 Journal of Practical Stomatology
关键词 RayPex5根尖定位仪 根管工作长度 准确性 Raypex5-apex locator Root canal working length Accuracy
  • 相关文献

参考文献5

二级参考文献17

  • 1贾淑娟,古向生.口腔X线检查放射防护的新进展[J].国外医学(口腔医学分册),2004,31(4):326-328. 被引量:11
  • 2韩晶,张运奎,谢鹏峰,姜涛,曾令芳.三种测量根管工作长度方法准确性的评价研究[J].山东大学基础医学院学报,2005,19(4):254-255. 被引量:6
  • 3张芳,王非,晋江丰,臧满丽.根尖片数字化X线摄影和传统X线摄影在根管治疗中应用比较[J].中日友好医院学报,2005,19(5):272-274. 被引量:7
  • 4王素文.根管充填程度对疗效的影响——附222例分析[J].中华口腔医学杂志,1982,1:33-33.
  • 5王晓仪.前牙根尖孔位置的临床意义[J].中华口腔科杂志,1985,20:164-164.
  • 6Weine FS. Calculation of working lenth. In:Endodontic therapy. 5th ed. St. Louis: Mosby, A Times Mirror Company, 1996.395 - 477
  • 7Yamashita Y. A sdudy of a new electric root canal measuring device using relative values of frequency response : Influence of the diameter of apical foramam, the size of electode, and the concentration of sodium hypochlorite. J Jpn Conserve Dent,1990,33 (3) :547
  • 8Seidberg BH, Alibrandi BV, Fine H. Clinical investigation of measuring working lengths of root canals with an electronic device and with digital-tactile sense. J Am Dent Assoc, 1975,90(2) :379
  • 9Marfis A,Sylaras SN,Georgopoulou M. Study of the apices of human permanent teeth with the use of a SEM. Oral Surg Oral Med Oral Pathol,1994,77(2):172
  • 10Sanderink C, CH,Huiskens R,vander stelt DF,et al. Image quality of direct digital intraoral x - ray senosors in assessing root canal length. Oral Sung Oral Med Oral Pathol, 1994,78 : 125.

共引文献19

同被引文献42

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部