摘要
目的:建立高效液相色谱-质谱联用法测定人血浆中兰索拉唑的浓度。方法:血浆样品经处理后,采用Waters AtlantisC_(18)色谱柱(100 mm×2.1 mm,3μm),流动相为甲醇-0.01%甲酸溶液(70:30),以电喷雾电离源(ESI)正离子检测,奥美拉唑为内标。结果:兰索拉唑血浆质量浓度的线性范围为5.0~1 500.0 ng·ml^(-1)(r=0.999 8),,提取回收率为78.2%~88.9%(n=5),方法回收率为90.3%~101.9%(n=5),日内和日间RSD均小于10%。结论:本法专属性强,样品处理方便,灵敏度高,适用于兰索拉唑药动学研究。
Objective: To establish an LC-MS/MS method for the determination of lansoprazole in the human plasma. Method: The lansoprazole was deproteined from plasma and the Waters Atlantis C18(100 mm x 2. 1 mm,3 μ m) column was used. The mobile phase consisted of methanol-0.01% formic acid solution(70: 30). Eleetrospray ionization (ESI) source was applied and operated in the positive mode. Omeprazole was used as the internal standard. Result: The calibration curve was linear within 5.0-1 500.0 ng. ml-1 ( r = 0.999 8 ). The extraction recovery was within 78.2% -88.9% ( n = 5 ) and the method recovery was within 90. 3% -101.9% ( n = 5 ). The relative standard deviation of within day and between-day were less than 10%. Conclusion: The method is proved to be ro- bust, convenient, sensitive and suitable for the clinical investigation of lansoprazole pharmacokinetics.
出处
《中国药师》
CAS
2010年第4期509-511,共3页
China Pharmacist
基金
盐城市科技发展项目(编号:YZ2007016)