Sysmex CA-1500全自动血凝分析仪混匀马达替换后性能评价

To Evaluate Performance of Sysmex CA-1500 Automatic Blood Coagulation Analyzer Mixing Motor Replaced
下载PDF
导出
摘要 目的评价用手机震动器替换Sysmex CA-1500全自动血凝分析仪混匀马达后的性能。方法测定混匀马达被替换后Sysmex CA-1500全自动血凝分析仪的PT、APTT、FIB三个项目,并与用Sysmex CA-500全自动血凝分析仪测定结果比对。结果PT、APTT、FIB三个项目测定的准确性、精密度及线性情况良好,结果可靠。且与Sysmex CA-500全自动血凝分析仪比对结果一致。结论用手机震动器替换血凝分析仪混匀马达,符合使用要求,延长仪器使用寿命。 Objective To evaluate Sysmex CA-1500 automatic blood coagulation analyzer performance after the motor was replaced by the cell phone vibrator.Methods After the replacement of mixing motor Sysmex CA-1500 automatic blood coagulation analyzer determines the PT,APTT,FIB to reflect the instrument accuracy,precision,linear determination,and with Sysmex CA-500 automatic blood coagulation analyzer results against.Results After the replacement of mixing motor Sysmex CA-1500 automatic blood coagulation analyzer for PT,APTT,FIB measured the accuracy,precision,and linear in good condition,results is reliable,with Sysmex CA-500 automatic blood coagulation analyzer consistent than the results.Conclusions Cell mobile phone vibrator can replace mixing motor of Sysmex CA-1500 automatic blood coagulation analyzer,The performance is good after the replacement of equipment,to meet the requirements,It can provide patients with timely and accurate services,and cost savings for hospitals.
出处 《生物医学工程学进展》 CAS 2010年第1期29-32,43,共5页 Progress in Biomedical Engineering
关键词 全自动血凝分析仪 PT、APTT、FIB 准确性 精密度 线性 automatic blood coagulation analyzer PT APTT FIB accuracy precision linear