期刊文献+

基于CT图像三维重建的颅内电极定位方法的研究 被引量:2

Method for Localizing the Subdural Electrodes Using 3-D Reconstruction of CT Imaging
下载PDF
导出
摘要 难治性癫痫的手术有时需要通过植入颅内电极进行脑电检测来确定致痫灶,为了准确定位致痫灶,医生需要知道电极的位置,但目前缺乏有效的电极定位方法。本文开发了一种利用植入电极后CT图像序列进行电极定位的方法。首先使用面绘制三维重建出电极序列图像,并对各个电极进行识别和标识,然后将面绘制重建的颅骨图像叠加到电极图像上,建立电极的位置信息与人体解剖结构的联系,最后实现颅内电极定位。经过对10套植入电极后的CT图像进行定位检测,证实基于CT图像三维重建的电极定位方法与目前已有的定位方法比较具有明显优势。 Intracranial EEG-monitoring with subdural electrodes is often used to localize the epileptic focus in epilepsy surgery. After implantation, it is imperative to know the position of subdural electrode in order to correlate the zone of ictal onset with precise anatomic structures. However, there are no efficient localization methods. We presented a method to localize the subdural electrodes using post-implanted CT image data. First, the surface of all of the electrodes and the skull were rendered. Then the user recognized and labeled individual electrodes. Finally we added the image of skull to the image of electrodes. The relation of electrode and anatomic structure was found. Ten sets post-implanted CT image data were used to evaluate this method' s efficacy. Compare to present methods, we obtained the conclusion that the method above was useful to localize the subdural electrodes.
作者 柳渊 严汉民
出处 《北京生物医学工程》 2010年第2期116-119,165,共5页 Beijing Biomedical Engineering
关键词 癫痫手术 颅内电极定位 面绘制 VTK epilepsy surgery subdural electrode surface rendering VTK
  • 相关文献

参考文献10

  • 1Engel Jr J,Wiebe S,French J,et al.Practice parameter:temporal lobe and localized neocortical resections for epilepsy[J].Epilepsia,2003,44:741-751.
  • 2杜建军,郝玮琳,刘有军,乔爱科.医学图像三维可视化系统框架设计与开发[J].北京生物医学工程,2008,27(6):569-573. 被引量:4
  • 3Tao JX,Susan HE,Maria Baldwin,et al.The accuracy and reliability of 3D CT/MRI co registration in planning epilepsy surgery[J].Clinical Neurophysiology,2009,120:748-753.
  • 4Morris K,O'Brien TJ,Cook MJ.A computer generated stereotactic "virtual subdural grid" to guide resective epilepsy surgery[J].AJNR Am J Neuroradiol,2004,25:77-83.
  • 5Sebastiano F,Gennaro GDi,Esposito V.A rapid and reliable procedure to localize subdural electrodes in cryos urgical evaluation of patients with drug resistant focal epilepsy[J].Clinical Neurophysiology,2006,117:341-347.
  • 6Kovalev D,Spreer J,Honegger J,et al.Rapid and fully automated visualization of subdural electrodes in the presurgical evaluation of epilepsy patients[J].AJNR Am J Neuroradiol,2005,26:1078-1083.
  • 7Dalal SS,Edwards E,Kirsch HE.Localization of neurosurgically implanted electrodes viaphotograph MRI radiograph coregistration[J].Journal of Neuroscience Methods,2008,174:106-115.
  • 8周律,张志广,高小榕,高上凯.MRI图像中颅内栅格电极定位算法的研究[J].北京生物医学工程,2007,26(4):337-341. 被引量:2
  • 9Davis LM,Spencer DD,Spencer SS,et al.MR imaging of implanted depth and subdural electrodes:is it safe?[J].Epilepsy Research,1999,35:95-98.
  • 10Mehran Mahvash,Roy Konig,Jorg Wellmer.Coregistration of digital photography of the human cortex and cranial magnetic resonance imaging for visualization of subdural electrodes in epilepsy surgery[J].Neurosurgery,2007,61:5340-5345.

二级参考文献12

  • 1赵明昌,田捷,薛健,朱珣,何晖光,吕科.医学影像处理与分析开发包MITK的设计与实现[J].软件学报,2005,16(4):485-495. 被引量:41
  • 2Schroeder W, Martin K, Lorensen B. The visualization toolkit: an object-oriented approach to 3D graphics. 3rd ed. Publishers: Kitware lnc, 2003.
  • 3Ibanez L, Schroeder W, Ng L, et al. The ITK Software Guide. Kitware Inc, 2005.
  • 4Papyrus toolkit. Digital Imaging Unit. Geneva University Hospital. http://www, expasy, ch/U1N/htm|l/projects/papyrus /papyrus. html, 2006.
  • 5Antoine Rosset, Luca Spadola, Osman Ratib. OsirX: An Open- Source Software for Navigating in Muhidimensonal DICOM images. Journal of Digital Imaging[ J ], 2004, 17 (3) : 205 - 216.
  • 6Stanimire Tomov, Robert Bennett, Michael McGuigan, et al. Application of interactive parallel visualization for commodity-based clusters using visualization APIs. Computer & Graphics [J] ,2004,28:173 -278.
  • 7Chabrerie A,Ozlen F,Nakajima S,et al.Three-dimensional Reconstruction and Surgical Navigation in Pediatric Epilepsy Surgery.MICCAI' 98,Proceedings,1998:74-83
  • 8Noordmans HJ,van Rijen PC,van Veelen CW,et al.Localization of Implanted EEG Electrodes in a Virtual-Reality Environment.Comput Aided Surg,2001,6:241-258
  • 9Skrinjar O,Duncan J.Automatic Extraction of Implanted Electrode Grids.MICCAI' 99,Proceedings,1999:990-997
  • 10Pozdin MO.Automated Extraction of Subdural Grid Electrodes from Post-Implant MRI Scans for Epilepsy Surgery.Thesis for Master's Degree.Georgia Institute of Technology,2004

共引文献4