期刊文献+

不同扫描条件下的小动物成像评估(英文)

EVALUATION OF SMALL ANIMAL IMAGING UNDER DIFFERENT SCAN CONDITIONS
下载PDF
导出
摘要 为了评估在不同高压、滤光片厚度和管电流条件下的小动物成像,开发了一套动物成像微型CT装置,并在此装置上进行了不同高压、滤光片厚度、球管电流和曝光时间条件下的老鼠成像。以蒙特卡罗软件包PENELOPE为基础,开发了模拟本装置使用的X射线球管光谱的模型。同时,基于Matlab平台开发了计算X射线束各种参数,包括平均能量、光强变化以及透射率变化等的软件平台。评估了各种影响微型CT成像的参数对信噪比(SDNR)的影响,并分析了成像参数选择与信噪比之间的关系。文中一系列针对小动物成像和图像质量评估的工具和方法可被广泛应用于小动物成像领域。 To evaluae small animal imaging with individual different high voltage, filter thickness and tube current, an animal X-ray micro-computed tomography (micro-CT) system based on panel detector is developed and a rat is scanned by using the system with individual high voltage, tube current, filter thickness, and exposure time. A model is presented based on the Monte Carlo code PENELOPE for generating the X-ray spectra of X-ray tube used in the micro-CT system. A platform developed based on Matlab allows for calculating beam quality parameters, including the average energy of X-ray beam, the change of transmition rate and the input X-ray fluence. The factors affecting the signal difference to noise ratio (SDNR) of micro-CT are investigated and the relationship between SDNR and scan combinations is analyzed. A series of tools and methods are developed for small animal imaging and imaging performance evaluation in the field of small animal imaging.
出处 《Transactions of Nanjing University of Aeronautics and Astronautics》 EI 2010年第1期27-33,共7页 南京航空航天大学学报(英文版)
基金 Supported by the National Natural Science Foundation of China (60672104,10527003) the Nation-al Basic Research Program of China ("973"Program)(2006CB705705) the Joint Research Foundation of Beijing Mu-nicipal Commission of Education (JD100010607)~~
关键词 成像技术 微型CT 图像质量 蒙特卡罗模拟 imaging techniques micro-computed tomography (CT) imaging performance Monte Carlo simulation
  • 相关文献

参考文献10

  • 1Chun I,Cho M,Lee S,et al.X-ray micro-tomogra- phy system for small-animal imaging with zoom-in imaging capability. Physics in Medicine and Biology . 2004
  • 2Lee S,Kim H,Chun I,et al.A flat-panel detector based micro-CT system:performance evaluation for small-animal imaging. Physics in Medicine and Biology . 2003
  • 3Sterpin E,Salvat F,Cravens R,et al.Monte Carlo simulation of helical tomotherapy with PENELOPE. Physics in Medicine and Biology . 2008
  • 4Siewerdsen J,Waese A,Moseley D,et al.Spektr: a computational tool for X-ray spectral analysis and imaging system optimization. Physics in Medicine and Biology . 2004
  • 5Salvat F,Fernandez-Varea J,Sempau J.PENE- LOPE-2006:a code system for Monte Carlo simu- lation of electron and photon transport. . 2006
  • 6Hubbell J,Seltzer S.Tables of X-ray mass attenua- tion coefficients and mass energy-absorption coeffi- cients. NIST20899 . 1999
  • 7Feldkamp LA,Davis LC,Kress JW.Practical cone-beam algorithm. Journal of the Optical Society of America A: Optics, Image Science and Vision . 1984
  • 8Kastis,GA,Furenlid LR,Wilson DW,et al.Compact CT/SPECT small-animal imaging system. IEEE Transactions on Nuclear Science . 2004
  • 9LLOVET X,SORBIER L,CAMPOS C S,et al.Monte-Carlo simulation of X-ray spectra genera-ted by kilo-electron-volt electrons. Journal of Applied Physics . 2003
  • 10A.Miceli,R.Thierry,M.Bettuzzi,et al.Comparison of simulated and measured spectra of an industrial 450 kV X-ray tube. NIM A . 2007