摘要
目的评价中央角膜厚度(central corneal thickness,CCT)校正前后Icare回弹式眼压计(Icare reboundto-nometer,Icare)、Goldmann压平眼压计(Goldmann applanation tonometer,GAT)和PASCAL动态轮廓眼压计(dynamic contour tonometer,DCT)眼压测量值的一致性。方法比较CCT对Icare、GAT和DCT三种眼压计的眼压测量值的影响程度,同时使用Bland-Altman分析对CCT校正前后的Icare、GAT和DCT任意两种眼压计的眼压测量值进行一致性分析。结果Icare、GAT和DCT三种眼压计的眼压测量值与CCT的相关系数分别为:0.341(P<0.001)和0.333(P<0.001)和0.032(P=0.699),CCT校正后任意两种眼压计之间的一致性界限均较CCT校正前有所缩窄。结论Icare、GAT和DCT三种眼压计的眼压测量值在未经CCT校正之前不能简单互换,但经CCT校正后Icare和DCT的眼压测量值可以互换。
Objective To evaluate the consistency of the intraocular pressure (lOP) values measured with Icare rebound tonometer (Icare), Goldmann applanation tonometer (GAT) and dynamic contour tonometer (DCT) before and after central corneal thickness (CCT) correction. Methods The effects of CCT on the IOP values measured with Icare, GAT and DCT were compared. The consistency among the IOP values was evaluated with Bland-Altman analysis. Results The correlation coefficients of CCT and IOP values measured by Icare, GAT and DCT were 0. 341 (P〈0. 001), 0. 333 (P〈0. 001) and 0. 032 (P = 0. 699) respectively. The 95% limits of the agreement between every two tonometers after CCT correction were narrower than those before correction. Conclusion The IOP values of the three tonometers can not be interchanged in clinic before CCT correction. However the IOP values of Icare and DCT can be exchanged after correcting according to CCT.
出处
《华南国防医学杂志》
CAS
2010年第2期100-103,106,共5页
Military Medical Journal of South China