期刊文献+

压电式测力平台系统输出压力中心参数的实验研究 被引量:4

Experimental Study on Center of Pressure Parameter for Output Pressure of Piezoelectric Force Plate System
下载PDF
导出
摘要 目的:探讨基于压电式测力平台检测压力中心算法的精确性,为其推广应用研究提供客观依据。方法:以压力中心算法推导为基础,应用DLUT6040-1压电式测力平台分别从静态与动态模式进行2次测试,每次测试时,分别有2位操作者重复测试2次,并对静态与动态测试结果进行分析。结果:2次测试结果各指标高度相关,动态验证与静态验证相比,COP输出力值的计算误差基本没有显著差异。结论:压电式测力平台用于评定人体压力中心具有较高的精确性,适用于各群体人体步态研究及矫形康复行走机能评定,可进一步推广使用。 Objective To study the reliability and accuracy of center of pressure measured by piezoelectric force plate, and to provide a useful reference for further application. Methods Piezoelectric force plate was utilized to measure the center of pressure twice a week. All the testing was performed by two operators, respectively, with dynamic and static measurement at each assessment in the study. Results There were moderate to high correlations in all the parameters of COP between the two assessments. Conclusion The COP measured by piezoelectric force plate has high reliability and accuracy. It can be successfully used in evaluation of gait and rehabilitation.[Chinese Medical Equipment Journal,2010,31 (4):21-23]
出处 《医疗卫生装备》 CAS 2010年第4期21-23,共3页 Chinese Medical Equipment Journal
基金 大连理工大学青年基金项目(893204)
关键词 静态 动态 压力中心 验证 static testing dynamic testing center of pressure validation
  • 相关文献

参考文献11

  • 1邹继豪,袁庆成.人体动态力的实验研究[M].大连:大连理工大学出版社,1990:37-38.
  • 2张贻恭,刘恩福.晶花传感器及其在大力值测量中的应用[J].辽宁工学院学报,1999,19(5):1-5. 被引量:1
  • 3Ekdahl C,Jarnlo G B,Andesson S I. Standing balance in healthy subjects[J]. Scand J Rehab Med, 1989,21 : 187-195.
  • 4Hall M G, Fleming H E, Dolan M J. Static in situ calibration of force plates[J]. Journal of Biomechanics, 1996,29:659-665.
  • 5Gill H S ,O'Connor J J. A new testing rig for force platform calibration and accuracy tests[J]. Gait & Posture, 1997,5 (3) :228-232.
  • 6Gustav H,Gautschi. Piezoelectric Multicomponent Force Transducers and Measuring Systems[M]. Switzerland : Kistler Company, 1978 : 27.
  • 7元文学,孟昭莉,刘海斌.多分量压电晶体测力平台系统静态标定方法的研究[C]//第十一届全国运动生物力学学术交流大会论文汇编.成都:运动生物力学分会,2006:136-137.
  • 8Winter D A. Human balance and posture control during standing and walking[J]. Gait & Posture, 1995,3(4) : 193-214.
  • 9Penelope J K, James J D. Accuracy of a Portable Force Plate in assessing force and Center of Pressure Estimates Under Static Loading [J]. Gait & Posture, 1997,5 (2) : 178-179.
  • 10Maarten F B, Henk C S. Accuracy of Determining the Point of Force Application with Piezoelectric Force Plates[J]. Journal Of Biomechanics, 1990,23(7) :705-710.

二级参考文献3

  • 1张贻恭.石英晶体力传感器的单晶(djm)转换及晶组排列与组合[J].传感技术学报,1988,1(1):18-25.
  • 2董显铨.压电式传感器设计中有关压电晶体若干考虑.全国振动计量交流会议[M].北京,1975..
  • 3刘恩福,吕希敏,张贻恭.积累式冲击扭矩测试仪研制[J].大连理工大学学报,1999,39(1):86-90. 被引量:3

共引文献1

同被引文献69

引证文献4

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部