钛合金假体,偏离和视域大小对骨密度的定量CT测定的影响(英文)

EFFECTS OF TITANIUM PROSTHESIS,OFFSET AND SIZE OF FIELD OF VIEW ON BMD MEASUREMENTS USING QUANTITATIVE COMPUTED TOMOGRAPHY
下载PDF
导出
摘要 为了评价用定量CT测定假体植入后周围骨密度变化这一方法的可行比性,本文用人皮质育和松质骨试样研究了钛合金假体,偏离和视域大小对骨密度的定量CT测定精度的影响、取自新鲜人尸体胫骨的14个皮质骨和14个松质骨试样(分别组成两列)用于本研究。骨试样分别在含有或不含有钛合金假体、同轴(骨试样列的纵轴线与CT扫描平面的轴线平行且重合)或偏离(骨试样列的纵轴线与CT扫描平面的轴线平行但偏离14厘米)、10厘米或30厘米视域条件下进行CT扫描和图象重构。实验结果表明:当骨试样列的纵轴线与CT扫描平面的轴线同轴时,由于钛合金假作的存在对皮质骨和松质骨BMD测量值的平均影响都小于1%,视域大小对骨密度测量的影响也很小。然而,当骨试样列中插入钛合金假体时,偏离会引起BMD测量值的显著不同(P<0.0001)。因此,在临床中用定量CT测定假体植入后周围骨密度变化时必须考虑到这一因素。 To investigate the accuracy of quantitative computed tomography (QCT) as a method to measure bone density(BMD) In the vicinity of a titanium prosthesis, the authors investigate the effects of(1) titanium prosthesis,(2) offset of the longitudinal axis of bone to be examined from that of the gantry of CT scanner,and(3) size of field of view, and the combine of these effects on these measurements by use of cadaveric bone specimens.
出处 《医用生物力学》 EI CAS CSCD 1998年第2期72-79,共8页 Journal of Medical Biomechanics
关键词 定量CT 骨密度 钛合金假体 偏离 视域大小 quantitative computed tomography,bone mineral density,titanium prosthesis,offset,size of field of view
  • 相关文献

参考文献12

  • 1McBroom RJ, Hayes WC, Edwards WT, et al. Prediction of vertebral body compressive fracture using quantitative computed tomography. J Bone Joint Surg 1985; 67A: 1206-1214.
  • 2Sartoris DJ, Andre M, Resnick C, et al. Trabecular bone density in the proximal femur:quantitative CT assessment. Radiology 1986; 160: 707-712.
  • 3Harris WH, Sledge CB. Total hip and total knee replacement (I). N Eng J Med 1990;323: 725-731.
  • 4Kilgus DJ, Shimaoka EE, Tipton JS, Eberle RW. Dual-energy X-ray absorptiometry measurement of bone mineral density around porous-coated cementless femoral implants. J Bone Joint Surg 1993, 75B: 279-287.
  • 5Cann CE. Quantitative CT for determination of bone mineral density: a review. Radiology 1988,166: 509-522.
  • 6Adolphson P, Sivers KV, Dalen N, et al. Bone and muscle mass after hip arthroplasty. Acta Orthop Scand 1993, 64: 181-184.
  • 7Robertson DD, Weiss PJ, Fishman EK, et al. Evalluation of CT techniques for reducing artifacts in the presence of metallic orthopedic implants. J Comput Assist Tomogr 1988; 12:236-241.
  • 8Markel MD, Morin RL, Roy RG et al. Effect of titanium endoprostheses on bone mineral density.measurements, using quantitative computed tomography. Am J Vet Res 1992; 53: 2105-2108.
  • 9Cacak R. Quality control for CT scanners. In: Seeram E, eds. Computed tomography, physical principles, clinical applications & quality control. Philadelphia: Saunders WB Press, 1994,232-250.
  • 10Feng Z, Ziv I, Rho J. The accuracy of computed tomography-based linear measurements of human femora and titanium stem. Invest Radiol 1996, 31: 333-337.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部