摘要 目的:探讨迈瑞5300血细胞分析仪的临床应用价值。方法按照国际血液学标准化委员会(ICSH)制定的评价标准对迈瑞5300血细胞的几项参数进行统计学分析。结果:100例患者检测质量控制良好,稳定性高,CV<5%;线性γ>0.999;与ICSH推荐的参考方法和Sysmex SF3000血细胞分析仪的对比,相关性均较好(γ>0.970);胆红素和甘油三酯对各参数基本无干扰(P>0.005);在室温条件下至少可稳定48h,不精密度的CV均<5%;WBC分类重复性测试和异常白细胞检测均较理想。结论:迈瑞5300血细胞分析仪校准方便,质量控制良好,稳定性高;线性检查理想;携带污染率非常低,对大批量的全血细胞计数(CBC)和白细胞分类计数(DC)能进行快速和有效的筛检。 Objective:To expore clinical application and analysis of mindray hematology analyzer BC-5300. Methods:To evaluating the 5300 according to the guidelines of international council for standardization in haematology (ICSH). Results :5300 had an excellent linearity for WBC,RBC,Hb and PLT (γ0.999) and stability and quality control (both CV5%). The correlation of CBC parameters between 5300 and ICSH reference methods or Sysmex SF3000 was excellent (γ0.970).Interfering study demonstrated minimal interference of high concentration of bilirubin or triglyceride for CBC(P0.05).The stability of blood samples for CBC parameters at room temperature (RT) was very stable for 48h,and imprecision was excellent for all CBC parameters(CV5%). Reperformance test of WBC and detection of abnormal leukocyte were perfect. Conclusion:5300 automated hematology analyzer with stability and fine quality control is calibrated conveniently. It also had excellent linearity and lower carryover so that it is rapid and efficiently to test CBC and DC in large quantities.
作者 张养民 白妮
出处 《陕西医学杂志》 CAS 2010年第1期101-103,共3页 Shaanxi Medical Journal
关键词 肺气肿/血液 血细胞计数 质量控制 Pulmonary emphysema/blood Blood cell count Quality control
  • 相关文献

参考文献4

  • 1Villanova PA.National committee for clinical laboratory standards leukocyte differential counting NCCLS approved standard H20-A2[M].Publisher:Prentice Hall,1992:1500-1507.
  • 2International council for standardization in hematology.Guidelines for evaluation of blood cell analysis including those used for differential leucocytes and reticulocyte counting and cell marker applications.Prepared by clinical laboratory hematology[J].Clin Lab Haem,1994,50(16):154-175.
  • 3廖益,王冬梅,金平,李晋蜀.Sysmex XT-2000i高端血细胞分析仪性能概况综合评价[J].实用医学杂志,2006,22(21):2528-2530. 被引量:4
  • 4熊立凡,李树仁.临床检验基础.第3版.北京:人民卫生出版社,2001:7-55.

二级参考文献3

  • 1束国防,高茂馗,芦慧霞.XE-2100血液分析仪的应用评价[J].现代医学,2005,33(5):323-325. 被引量:5
  • 2叶应妩;王毓三.全国临床检验操作规程,2000.
  • 3International council for Standardization in Hematology..Guidelines for evaluation of blood cell analysis including those used for differential leucocytes and reticulocyte counting and cell marker applications.Prepared by the ICSH expert panel on cytometry.