MRI监控高强度聚焦超声辐照升温的有效性和准确性研究 被引量:3

Efficacy and accuracy study of MRI thermometry monitor in high intensity focused ultrasound exposure
下载PDF
导出
摘要 目的探讨高强度聚焦超声(HIFU)辐照过程中,MRI靶区温度监控的有效性和准确性。方法 HIFU辐照离体牛肝组织,对比MRI的温度图(T-Map)显示坏死区域与实际解剖坏死区域面积,以检验T-Map监控的有效性;HIFU辐照仿组织透明体模,对比MRI所测温度与光纤传感器所测温度。结果 T-Map所示的坏死区面积和实际坏死区面积出现统计学差异(P>0.05)的概率约为25%;对小于1℃/s的缓慢温度变化,MRI和光纤的测温结果较为一致,二者的平均差异在2.6℃以内;而对于剧烈的温度变化,二者的平均差异达到5℃以上。结论 MRI的T-Map可对靶区坏死情况提供一定的参考信息,但结果并不完全可靠,同时MRI测温的精确性还有待进一步加强。 ABSTRACT Objective To investigate the accuracy and efficacy of MRI thermometry during high intensity focused ultrasound (HIFU) exposure. Methods HIFU was applied on bovine liver tissue in vitro, and then the necrosis square of MRI T-Map and practical tissue was compared, HIFU was also applied to expose the tissue-mimicking phantom, and then the temperature measured by MRI and optical fiber temperature sensor was compared to verify the accuracy of MRI thermometry. Results The probability for necrosis square of T-Map statistically different from the necrosis square of the practical tissue was 25% (P 〉 0.05); When the temperature change was less than 1℃/s, the result of MRI thermometry was comparatively consistent with the optical fiber sensor, their mean difference was less than 2.6℃ , while for the rapid temperature change, the mean difference was more than 5.0℃. Conclusion T-map of MRI can provide some information for necrosis of target area, but the result is not reliable enough, and the accuracy of MRI thermometry is need to be improved.
出处 《临床超声医学杂志》 2010年第3期148-151,共4页 Journal of Clinical Ultrasound in Medicine
基金 国家科技支撑计划(2006BAI 03A02) 重庆市科委重大科技专项(2007AA5020) 重庆医科大学科技基金(QD200542)
关键词 MRI测温 高强度聚焦超声 光纤温度传感器 Magnetic resonance thermometry High intensity focused ultrasound Optical fiber temperature sensor
  • 相关文献

参考文献10

二级参考文献68

  • 1耿晓鸣,王信红,石珍龙,宋文虎.高强度聚焦超声肿瘤治疗系统专项技术研究[J].声学技术,2004,23(F11):93-97. 被引量:3
  • 2Christakis D M, Pavlou O, Velev K, et al. High intensity focused ultrasound ablation of kidney guided by MRI[J]. Ultrasound Med Bio, 2004, 30(3): 397-404.
  • 3Lynn J G, Zwemer R L, Chick A J, et al. A new method for the generation and use of focused ultrasound in experimental biology[J]. J Gen Physiol, 1942,26(2): 179-193.
  • 4Fry W J, Mosberg W H, Barnard J W, et al. Production of focal destructive lesions in the central nervous system with ultrasound[J]. J Neurosurg, 1954, 11(5):471-478.
  • 5Clarker R L, Ter Haar G R. Temperature rise recorded during lesion formation by high-intensity focused ultrasound[J]. Ultrasound Med Bio, 1997, 23(2): 299-306.
  • 6Bihrle R, Foster R S, Sanghvi N T, et al. High intensity focused ultrasound for the treatment of benign prostatic hyperplasia: early United States clinical experience[J]. J Urol, 1994, 151(5): 1271-1275.
  • 7Vallancien G, Charter KE, Bataile N, et al. Focused extracorporeal pyrotherapy[J]. Eur Urol, 1993, 23(Supl 1):48-52.
  • 8Solomon S B, Nicol T L, Chan D Y, et al. Histologic evolution of high intensity focused ultrasound in rabbit muscle.[J]. Invest Radiol, 2003, 38(5): 293-301.
  • 9Malcolm A L, Ter Haar G R Ablation of tissue volumes using high intensity focused ultrasound[J]. Ultrasound Med Biol, 1996, 22(5): 659-669.
  • 10Seip R, Ebbini E S. Noninvasive estimation of tissue temperature response to heating fields using diagnostic ultrasound[J]. IEEE Trans Biomed Eng, 1995, 42(8):828-839.

共引文献17

同被引文献31

  • 1肖子文,肖雁冰,刘丽萍,李发琪,王智彪.高强度聚焦超声损伤兔肝脏的剂量及二维超声实时监控[J].重庆医科大学学报,2005,30(1):42-45. 被引量:1
  • 2刘春梅,陈锦云,柯丹,徐杰,雷令,李发琪,王智彪.工作周期对脉冲高强度聚焦超声辐照离体牛肝的影响[J].中国超声医学杂志,2007,23(8):567-569. 被引量:8
  • 3Fruehatif JH. Hack W. Eierm nnn A,et al. High intensity focused ultrasound for ihe targeted deslruction of uterine tissues:experiences from a pilot study using a mobile HIFLJ unit. ArchGynecol Obstet,2008,227(2) : 143-150.
  • 4Blana A, Rogenhofer S, Ganzer R, el al. Eight years' experiencewith high intensity focused ultrasonography for treatment of lo-calized prostale cancer. Urology, 2008, 72(6) : 1329-1333.
  • 5Wu F. ter Haar G,Chen WR. High-intcnsily focused ultrasoundablation of breast cancer. Expert Rev Anlicancer Ther. 2007,7(6): 823-831.
  • 6Li JJ . G ii M F. Luo G Y, et al. Complicalions of high intensityfocused ultrasound for patients wilh hepatocellular carcinoma.Technol Cancer Res Treat, 2009 , 8(3) : 217-224.
  • 7Yang FY, Chiu WH,Liu SH , et al. Functional changes in arter-ies induced by pulsed high intensity focused ullrasound. IEEETrans Ultrason Ferroeleclr Freq Control. 2009,56(12): 2643-2649.
  • 8Burke CW, Klibanov AL, Sheehan JP, et al. Inhibition of gliomagrowth by microhuhble activation in a subcutaneous model usinglow duty cycle uhrasound without significant heating. J Neuro-surg. 2011, 111(6): 1654 1661.
  • 9Wu. F. , Wang. Z. B. . Chen. W. Z. , Bai, J. ? Zhu. H. andQiao.T. Y. : Preliminary experience using high intensity focusedultrasound for ihe treatment of patients wilh advanced stage renalmalignancy. J Urol. 2003.170:2237.
  • 10Tran. B. C.,Seo, J. , Hall. T. L. , Fowlkes. J. R and Cain.C.A.:Microlmhble-enhanced cavitation for noninvasive ultrasound surgery. IEEE Trans Ultrason FVrroelectr Freq Control.2003,50:1296.

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部