Frequency Response of the Sample Vibration Mode in Scanning Probe Acoustic Microscope 被引量:2

Frequency Response of the Sample Vibration Mode in Scanning Probe Acoustic Microscope
下载PDF
导出
摘要 Based on the interaction mechanism between tip and sample in the contact mode of a scanning probe acoustic microscope (SPAM), an active mass of the sample is introduced in the mass-spring model. The tip motion and frequency response of the sample vibration mode in the SPAM are calculated by the Lagrange equation with dissipation function. For the silicon tip and glass assemblage in the SPAM the frequency response is simulated and it is in agreement with the experimental result. The living myoblast cells on the glass slide are imaged at resonance frequencies of the SPAM system, which are 20 kHz, 30 kHz and 120 kHz. It is shown that good contrast of SPAM images could be obtained when the system is operated at the resonance frequencies of the system in high and low-frequency regions. Based on the interaction mechanism between tip and sample in the contact mode of a scanning probe acoustic microscope (SPAM), an active mass of the sample is introduced in the mass-spring model. The tip motion and frequency response of the sample vibration mode in the SPAM are calculated by the Lagrange equation with dissipation function. For the silicon tip and glass assemblage in the SPAM the frequency response is simulated and it is in agreement with the experimental result. The living myoblast cells on the glass slide are imaged at resonance frequencies of the SPAM system, which are 20 kHz, 30 kHz and 120 kHz. It is shown that good contrast of SPAM images could be obtained when the system is operated at the resonance frequencies of the system in high and low-frequency regions.
机构地区 Institute of Acoustics
出处 《Chinese Physics Letters》 SCIE CAS CSCD 2010年第5期178-181,共4页 中国物理快报(英文版)
基金 Supported by the National Natural Science Foundation of China under Grant Nos 10774113 and 10834009, and the Doctoral Fund of Ministry of Education of China under Grant No 20070247042.
  • 相关文献

参考文献22

  • 1Kolosov O V, Castell M R, Marsh C D and Briggs G A D 1998 Phys. Rev. Lett. 81 1046.
  • 2Striegler A, Kohler B and Bendjus B 2007 Proc. SPIE6528 65281B.
  • 3Passeri D, Rossi M, Alippi A, Bettucci A, Manno D, Serra A, Filippo E, Lucci M and Davoli I 2008 Superlattices Mi- crostruct. 44 641.
  • 4Yin Q R, Yu H F, Zeng H R, Li G R and Xu Z K 2005 Mater. Sci. Eng. B 120 100.
  • 5Zhao K Y, Zeng H R, Song H Z, Hui S X, Li G R, Yin Q R, Shimamura K, Kannan C V, Villora E A G, Takekawa S and Kitamura K 2008 Chin. Phys. Lett. 25 3429.
  • 6Diebold A C 2005 Science. 310 61.
  • 7Shekhawat C S and Dravid V P 2005 Science. 310 89.
  • 8Tittmann B R, Du J, Ebert A and Piccione B 2005 Proc. SPIE 5768 11.
  • 9Ebert A, Tittmann B R, Du J and Scheuchenzuber W 2006 Ultrasound Med. Biol. 32 1687.
  • 10Banerjee S, Gayathri N, Shannigrahi S R, Dash S, Tyagi A K and Raj B 2007 J. Phys. D: Appl. Phys. 40 2539.