非接触式眼压计与压陷式眼压计的临床应用效果分析 被引量:6

Analysis in clinical efficiency of non - contact tonometer and indentation tonometer
原文传递
导出
摘要 目的对比非接触式眼压计与压陷式眼压计所测量的眼压值,以评价非接触式眼压计的临床应用价值。方法将在门诊做眼压筛查的160例患者320只眼纳入本观察,采用自身前后对照法,对比非接触式眼压计(简称NCT)与压陷式眼压计测量的眼压值。结果用NCT测量的眼压值显著低于压陷式眼压计的测量值。结论NCT测量的眼压值显著低于压陷式眼压计测量的眼压值,对于NCT眼压在2.67kPa者,最好用压陷式眼压计再测1次或间隔复诊。 Objective To compare the measured IOP values by non-contact tonometerand indentation tonometer in order to evaluate the non-contact tonometer in clinical application. Methods 320 eyes of 160 patients for screening of lOP in the clinic were included in the observation, by self-control study, the IOP value of the non-contact tonometer (abbreviated NCT) and the indentation tonometer was compared. Results IOP values measured by NCT was lower than that measured by indentation tonometer. Conclusions IOP values measured by NCT was lower than that measured by indentation tonometer, for IOP value around 2.67 kPa measured by NCT, it is best to use indentation tonometer to retest or coming again for subsequent visit.
出处 《中国实用护理杂志》 北大核心 2010年第5期50-51,共2页 Chinese Journal of Practical Nursing
关键词 压陷式眼压计 非接触式眼压计 眼压 Indentation tonometer , Non-contact tonometer IOP
  • 相关文献

参考文献5

二级参考文献12

  • 1张宇燕,孙兴怀.常见眼压计的临床评价[J].眼视光学杂志,2004,6(2):132-135. 被引量:31
  • 2阮丽萍,童素莲.心理干预对非接触式眼压计眼压测量值的影响[J].现代中西医结合杂志,2007,16(6):819-820. 被引量:7
  • 3胡茂生 刘淳诗.非接触式眼压计与Goldmann氏压平眼压计对比研究[J].中华眼科杂志,1985,:111-111.
  • 4Pandav SS A, Gupta A. Reliability d Pro Ton and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984.
  • 5ShieldsMB.Thenon-contacttonometer.Itsvalueandlimitations[J].SurvOphFthalmol,1980,24:211.
  • 6施益浩.非接触眼压计正常眼眼压测定[J].中华眼科杂志,.
  • 781,17:211-213.
  • 8EhlersN,BramsenT,SperlingS.Applanationtonometryandcentralcornealthickness[J].ActaOphthalmol,1975,53:34-43.
  • 9JohnsonM,KassMA,MosesRA,etal.Increasedcornealthi-cknesssimulatingelevatedintraocularpressure[J].ArchOphthalmol,1978,96:664.
  • 10WhitacreMM,SteinRA,HassaneinK.Theeffectofcornealthicknessonapplanationtonometry[J].AmJOphthalmol,1993,115:592-596.

共引文献38

同被引文献35

引证文献6

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部