期刊文献+

氢化物原子荧光法测定水中痕量砷的最佳实验条件研究 被引量:3

Study on Optimal Experimental Conditions of Hydride Generation Atomic Fluorescence Spectrum on Determination of Arsenic in Water
下载PDF
导出
摘要 文章通过对负高压、灯电流、载气流量、屏蔽气流量等仪器工作参数及硼氧化钾、硫脲-抗坏血酸添加量等工作条件的研究,建立了氢化物发生原子荧光法测水中砷的最佳工作条件。在选定的最佳条件下,方法的检出限为0.019μg/L。该方法操作简单、灵敏度高、准确度高,在实际工作中具有重要的指导意义。 In the paper, operational parameter of instrument such as minus high, voltage currents of lamp, carrier gas flow rate, and shielding gas flow rate which affected the test were optimized. Effects of KBH4 concentration and thiourea-ascorbic acid content on measurement results were studied. The optimal experimental conditions of determination of Arseni in water through hydride generation atomic fluorescence spectrometry were determined. Under optimal conditions, the detection limit was 0.019 μg/L. The method was simple, sensitive and accurate, which had important guiding meaning.
出处 《广东化工》 CAS 2010年第5期212-214,共3页 Guangdong Chemical Industry
关键词 氢化物发生原子荧光法 最佳试验条件 hydridegeneratinnatomicfluoresceneespectrometry water Arsenic optimalexperimentalconditions
  • 相关文献

参考文献3

二级参考文献7

  • 1梅树珍,刘爱萍,李智文,付云,刘长英,张桂兰,赵福生,马涛,林亮,王兰英,白玉平,杨德清,张志明,张文祥,刘艳春,田英普,胡志忠,银秀梅,闫涛,道尔金,张茂林.内蒙土右旗木头湖村慢性砷中毒流行病学调查报告[J].中国地方病学杂志,1996,15(4):237-240. 被引量:9
  • 2李树猷.氯化铁混凝法饮水除砷研究[J].中国供水卫生,1993,2:3-7.
  • 3庞进法.水中砷测定方法的初步改进[J].中国供水卫生,1997,3:26-26.
  • 4王晓.氧化物原子分光光度法测定水中砷[J].中国供水卫生,1997,1:30-30.
  • 5刘素君.VA—90气态原子化装置—原子吸收法测定水中砷[J].中国供水卫生,1996,1:9-9.
  • 6李树猷.饮用水中砷的危害及除砷措施.中国供水卫生,1993,(1):43-43.
  • 7胡家英,丁刚.双道氢化物发生原子荧光法同时测定水中砷和汞[J].中国供水卫生,2000,8(2):16-19. 被引量:1

共引文献74

同被引文献23

引证文献3

二级引证文献13

相关作者