期刊文献+

用于浅表肿瘤热疗的二元聚焦超声换能器及其热场分布 被引量:3

A DOUBLE ELEMENT FOCUSED ULTRASONIC TRANSDUCER AND ITS HEAT DISTRIBUTION USED IN HYPERTHERMIC ONCOLOGY FOR SUPERFICIAL TUMOR
下载PDF
导出
摘要 介绍一种由两个压电圆片振子(直径25mm)构成前向聚焦声场(焦距约为114mm)的超声波加热装置。体模实验测量表明:经该装置持续加热(超声功率约2×40W)10~15分钟后可在体模内达稳定温度分布,使换能器中心附近6mm×6mm×10mm范围内达41~45℃有效热疗高温(起始环境温度为20℃时),较适用于浅表肿瘤的加热治疗。 In this paper, a double element focused ultrasonic transducer and its heat distribution is presented. It is designed for using in hyperthermic oncology for superficial tumor. The transducer is consisted of two circular disk piezoelectricoscillators with same diameter of 25 mm and generats a focused ultrasonic field with a focus of 114 mm. It is performed by phantom experament tnat a 41~45 ℃ temperature distribution which fits to the need of hyperthermic oncology in the phantom near thecenter point part (about 6 mm×6 mm×10 mm volume) out of the transducer can be produced after heating for 10~15 minutes continunously with ultrasonic power about 2×40 Watts when the started temperature is about 20 ℃.
出处 《中国医学物理学杂志》 CSCD 1999年第1期39-40,共2页 Chinese Journal of Medical Physics
基金 天津市21世纪青年科学基金
关键词 二元聚焦 热场 超声波 换能器 肿瘤 热疗法 Ultrasonic hyperthermic oncology Double element focus Heat distribution
  • 相关文献

参考文献3

二级参考文献1

共引文献11

同被引文献14

  • 1林世寅,胡自省,毛慧生,张天泽.第七届国际肿瘤热疗会议概述[J].中国肿瘤临床,1996,23(9):671-675. 被引量:11
  • 2杨悦,Die Angewandte Makromolekulare Chimie,1999年,18卷,2期,70页
  • 3王智标,中国超声医学杂志,1999年,15卷,5期,241页
  • 4冯若,中华超声影像学杂志,1999年,15卷,3期,239页
  • 5户田康永 山路雅人 藤井麻美子.强力超音波照射にょる癌治にⅢする理諭解析とフアントム加温實驗[J].医用電子と生体工学,1998,12(6):15-21.
  • 6Lee R J, Buchanan M, Kleine LJ, et al. Arrays of multielement ultrasound applicators for interstitial hyperthermia. IEEE Trans BiomedEng, 1999 , 46 (7): 880-90
  • 7Lin WL, Yen JY, Chen YY, et al. Specific absorption rate ratio patterns of cylinder ultrasound transducers for breast tumor. Med Phy, 1998, 25 (6): 1041- 1048
  • 8Santacroce L,Minervini MM, Gaglirdi S, et al. Virtual reality: a simple joke or therapeutic instrument. Clin Ter, 1999, 150 (2):127-34
  • 9McGough, Robert J. Treatment planning for hyperthermia with ultrasound phased arrays. IEEE Trans Ultrason Ferroelect and Freq Control, 1996, 43 (6): 1074 - 1084
  • 10Ming D,Wan BK, Zhang LX, et al. Theoretical modeling study of the necrotic field during high-intensity focused ultrasound surgery.MedSci Monit, 2004, 10 (2): 19-23

引证文献3

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部