期刊文献+

低剂量CT扫描在鼻窦检查中的应用价值 被引量:7

Applied Value of Low-dose CT Scanning in Nasal Sinus Examination
下载PDF
导出
摘要 目的评价低剂量CT扫描在鼻窦检查中的临床应用价值。方法 100例患者分成少年组(5~15岁)和成年组(16~71岁),各50例,征得其同意在鼻窦常规剂量(150mAs)CT扫描后再行选定层面低剂量单层对照冠状位CT扫描,少年组mAs设为40mAs、25mAs,成年组设为50mAs、30mAs,其他扫描参数不变。由3位CT主治医师采用盲法评价CT图像,分为正常图像、图像有少许伪影、图像有严重伪影等三级对每帧图像进行质量评价。结果少年组和成年组低剂量CT扫描的单次扫描加权CT剂量指数(CTDIw)明显低于常规剂量扫描(P<0.01),少年组40mAs和成年组50mAs剂量扫描所得图像与常规剂量图像质量比较无明显下降,而少年组25mAs和成年组30mAs扫描图像与常规剂量图像质量比较差异有统计学意义。结论分别利用40mAs和50mAs低剂量对青少年和成年人进行鼻窦CT扫描,图像不影响诊断,可以有效降低患者的受辐射剂量。 Objective To assess the value of low dose CT scanning applied in nasal sinus examination. Methods 100 cases were divided into two groups, including youth group and adult group,with 50 cases for each group. After scanned by standard dose(150mAs) , all the patients were scanned with low dose. Those in youth group were scanned with 40mAs and 25mAs, and the others in adult group were scanned with 50mAs and 30mAs. CT images were evaluated by three doctors using blind method. The image quality was evaluated according to 3 grades:normal image,image with mild artifact,and image with serious artifact and the results were analyzed statistically. Results The CTDIw of low - dose CT scanning was obviously lower than that of standard dose ( P 〈 0.01 ). Compared with the images abtained from standard dose CT scanning,there was no significant difference in the quality of the images abtained from the dose of 40mAs for youth group and 50mAs for adult group,which obvious difference could be found in the images abtained from the dose of 25mAs for youth group and 30mAs for adult group. Conclusion CT scanning with the dose of 40mAs for adolescent and 50mAs for adults in nasal sinus examination Could significantly reduce the patient's radiation dose. The images could meet the needs of diagnosis.
出处 《医学研究杂志》 2010年第5期84-86,共3页 Journal of Medical Research
基金 广东省东莞市科技计划医疗卫生类科研项目(2008105150118)
关键词 鼻窦 体层摄影术 X线计算机 低剂量 Nasal sinus Tomography, X - ray computed Low - dose
  • 相关文献

参考文献6

  • 1Hein E, Rogalla P, Klingebiel R,et al. Low -dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol,2002,12(7) :1693 -1696.
  • 2任庆云,何杰,王大军,董玉龙.副鼻窦低剂量CT扫描的临床研究[J].中国辐射卫生,2005,14(2):119-119. 被引量:21
  • 3Sohaib SA,Peppercorn PD, Horrocks JA, et al. The effect of decreasing mAs on image quality and patient dose in sinus CT. Br J Radiol,2001, 74:157 - 161.
  • 4Tack D, Viviane De, Maertelaer, et al. Dose reduction in muhidetector CT using attenuation - based online tube current modulation. A JR, 2003,181(8) :331 -334.
  • 5Dinkel HP,Sonnenschein M, Hoppe H, et al. Low - dose multislice CT of the thorax in follow - up of malignat lymphoma and extrapulmonary primary tumors. Eur Radiol, 2003,13 ( 6 ) : 1241 - 1249.
  • 6王新怡,李福生,杨志强,张经建,马振申.多层螺旋CT低剂量扫描在副鼻窦检查中的应用[J].山东大学耳鼻喉眼学报,2009,23(2):59-62. 被引量:11

二级参考文献17

  • 1Matthias H, Amir A, Riva R, et al. Multidetector CT of the paranasal sinus: potential for radiation dose eduction[J]. Radiology, 2007, 243(3):847-852.
  • 2Mulkens T, Broers C, Fieuws S, et al. Comparison of effective doses for low-dose MDCT and radiographic examination of sinuses in children[J]. AIR, 2005, 184:1611-1618.
  • 3Sohaib S A, Peppercorn P D, Horrocks J A, et al. The effect of decreasing mAs on image quality and patient dose in sinus CT[J]. Br J Radiol, 2001, 74:157-161.
  • 4Zammit-Maempel I, Chadwick C L, Willis S P. Radiation dose to the lens of eye and thyroid gland in paranasal sinus multislice CT[J].Br J Radiol, 2003, 76(906) :418-420.
  • 5Hojreh A, Czemy C, Kainberger F. Dose classification scheme for computed tomography of the paranasal sinuses[ J ]. Eur J Radiol, 2005, 56(1):31-37.
  • 6Marmolya G, Wiesen E J, Yagan R, et al. Paranasal sinuses: low-dose CT[J]. Radiology, 1991, 181:689-691.
  • 7Bernhardt T M, Rapp-Bemhardt U, Messel A, et al. CT scaning of the paransal sinuses axial helical CT with reconstrusction in the coronal direction versus coromal helical CT[ J]. Br J Radiol, 1998, 71:846-851.
  • 8Shankar L, Evans K, hawke M, et al. An atlas of imaging of the paranasal sinuses[J]. Dunitz, London, 25.
  • 9Sohaib SA, Horrocks JA, Keene MA et al. The effect of decreasing mAs on image quality and patient dose in sinus CT[J]. The British journal of Radiology 2001,74:157 - 161.
  • 10Widberger JE, Mahnken AH, Schmitz-rodle T, et al. Individually adapted examination protocols for reduction of radiation exposure in chest CT[J]. Investigative Radiology 2001, 36(10):601- 611.

共引文献30

同被引文献83

引证文献7

二级引证文献63

相关作者

内容加载中请稍等...