回弹式眼压计与非接触眼压计测量眼压结果比较 被引量:1

The comparison of iCare rebound tonometer and non-contact tonometer
原文传递
导出
摘要 目的 比较应用爱科回弹式眼压计(iCare Rebound Tonometer,RBT)与非接触眼压计(non-contact tonometer,NCT)测量正常人群眼压结果的差异,并讨论RBT在眼科临床中的应用.方法 对131例(262只眼)分别用RBT和NCT在同一体位进行眼内压(intraocular pressure,IOP)的测量,对所得数据进行配对t检验及相关性分析.结果 NCT所测眼压平均值为(13.69±6.01)mmHg,RBT所测眼压平均值为(14.05±4.09)mmHg,NCT和RBT测得的眼压值无显著差异(P=0.23),两种眼压测量方式具有良好的相关性(相关系数0.56).结论 在进行正常人群高眼压筛选中,RBT和NCT测得眼压值无显著差异,有良好的相关性,RBT在进行青光眼的初步筛选中比NCT更为方便、快捷,值得推广. Objective To evaluate the clinical application of iCare rebound tonometer (iCare RBT) and compare with non-contact tonometer (NCT). Methods The intraocular pressures of 131 patients (262 eyes) were measured with iCare RBT and NCT respectively.The mean IOP measurement by RBT was compared with the measurement by NCT. Student's t-test was performed.Besults The mean IOP measured by NCT was 13.69± 6.01 mmHg and the mean IOP measured by RBT was 14.05± 4.09, there was no statistically significant difference between the mean IOP measured either by RBT or by NCT (P=0.23 ).It showed a good correlation between the two measurements (correlation coefficient 0.56).Conclusions The mean IOP is not statistically significant measured by RBT and NCT.Its ease to use, portability, and sterility makes RBT an attractive tonometer.
出处 《中国实用眼科杂志》 CSCD 北大核心 2010年第5期503-505,共3页 Chinese Journal of Practical Ophthalmology
关键词 眼内压 回弹式眼压计 非接触眼压计 Intraocular pressure Rebound tonometer Non-contact tonometer
  • 相关文献

参考文献11

  • 1Caprioli L,Speath GL.Comparison of visual field defects in the lowten-sion glaucomas with those in the high-tension glaucomas[J].Am J Oph-thalmol,1984,97:730-737.
  • 2The Advanced Glaucoma Intervention Study Investigators.The Advanced Glaucoma Intervention Study (AGIS).Ⅶ:the relationship between control of intraocular pressure and visual field deterioration[J].Am J Ophthalmol,2000,130:429-440.
  • 3Cllaborative Normal-Tension Glaucoma Study Group.Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures[J].Am J Ophthalmol,1998,126:487-497.
  • 4甄毅,王宁利,郭彦,张馨蓓.不同眼压计类型和测量体位对青光眼眼压测量值的影响[J].眼科研究,2009,27(1):55-58. 被引量:16
  • 5Vernon SA.Intra-eye pressurerange and pulse profiles in normals with the Pulsair[J].Eye,1993,7(Ptl):134-137.
  • 6张宇燕,孙兴怀.常见眼压计的临床评价[J].眼视光学杂志,2004,6(2):132-135. 被引量:31
  • 7方素珍,童素莲.护理干预在非接触式眼压计眼压测量中应用效果评价[J].现代中西医结合杂志,2007,16(17):2436-2437. 被引量:1
  • 8石晶明,蒋幼芹.非接触眼压计测量的评价——与Goldmann眼压计的比较[J].中国实用眼科杂志,2002,20(5):370-372. 被引量:28
  • 9KNIESTEDT C,LIN S,CHOE J,et al.Glinical comparison of contour and applanation tonometry and their relationship to pachymetry[J].Arch Ophthalmol,2005,123(11):1523-1537.
  • 10KANNGIESSER H E,KNIESTEDT C,ROBERT V C.Dynamic contour tonometry:presentation of a new tonometer[J].J Glaucoma,2005,14 (5):344-350.

二级参考文献46

  • 1张宇燕,孙兴怀.常见眼压计的临床评价[J].眼视光学杂志,2004,6(2):132-135. 被引量:31
  • 2Hara T.,Tsuru T.,王大江(译).通过昼夜体位变化的再现研究睡眠中眼压峰值的增加[J].世界核心医学期刊文摘(眼科学分册),2006,2(6):19-19. 被引量:7
  • 3Pandav SS,Sharma A,Gupta A,et al.Reliability of ProTon and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984.
  • 4Garzoz HJ,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy a comparison of three different tonometers[J].Cornea,2001,20(1):33-36.
  • 5Duch S,Serra A,Castanera J,et al.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma,2001,10(4):261-265.
  • 6Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794.
  • 7Munoz LA,Perez AL,Rahhal MS,et al.Assessment of applanation tonometry after hyperopic laser in situ keratomileusis[J].Cornea,2002,21(2):156-160.
  • 8Copt RP,Tomas R,Mermoud A.Corneal thickness and normal tension glaucoma[J].Arch Ophthalmol,1999,117(1):14-16.
  • 9Dekker PW,Robert YCA,Kanngiesser H,et al.Principles of contact lens tonometry[J].International Ophthalmology,1998,22(2):105-111.
  • 10Shah S,Chatterjee A,Mathai M,et al.Relationship between corneal thickness and messured intraocular pressure in a general opthalmology clinic[J].Ophthalmology,1999,106(11):2154-2160.

共引文献71

同被引文献3

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部