Raypex 5根尖定位仪测定乳牙根管工作长度准确性的实验研究 被引量:1

In vitro measurement accuracy of Raypex 5 electronic apex Iocator in primary teeth
下载PDF
导出
摘要 目的探讨Raypex 5根尖定位仪测定乳牙根管工作长度的准确性。方法将90颗离体乳牙(共144个根管)根据有无根尖吸收分为甲组(牙根无吸收,70个根管)及乙组(牙根有吸收,74个根管)。将两组牙齿放置于测量槽中,用Raypex 5测量根管工作长度,与肉眼下测量的根管实际工作长度进行比较。结果 Raypex 5根尖定位仪测量甲组的准确率为95.7%,测量乙组的准确率为91.9%,两组准确率的差异无统计学意义(P>0.05)。结论使用Raypex 5根尖定位仪能有效测量乳牙工作长度。 Objective: To evaluate the accuracy of Raypex 5 electronic apex locator in determining the working length of primary teeth.Methods: Ninty sound extracted primary teeth(144 total root canals,including 74 canals with visible resorption and 70 canals without) were selected.The teeth were embedded in an alginate model and electronic length were determined with Raypex 5 apex locator.Root canal lengths were measured visually to judge the accuracy of electronic reading.Results: The accuracy of Raypex 5 electronic apex locator in measuring the working length of primary teeth was 95.7% in group A and 91.9% in group B(P0.05).Conclusion: Raypex 5 is useful in determining the working length in primary teeth.
作者 赵文华
出处 《泰山医学院学报》 CAS 2010年第3期202-204,共3页 Journal of Taishan Medical College
关键词 根尖定位仪 工作长度 乳牙 electronic apex locator working length primary teeth
  • 相关文献

参考文献11

二级参考文献34

  • 1周艳,黄新文,金艳,邓莉华.ROOT ZX根尖定位仪在乳牙根管测量中的应用[J].临床口腔医学杂志,2005,21(6):341-342. 被引量:7
  • 2朱习华,董红,龙明生.ROOT ZX根管测量仪测量根管长度的临床分析[J].北京口腔医学,2006,14(1):63-63. 被引量:7
  • 3Goldberg F,De Silvio AC, Manfre S,et al. In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption[J]. J Endodont,2002:28(6) :461 - 463.
  • 4ElAyouti A, Weiger R, Lost C. The ability of Root ZX apexlocator to reduce the frequency of overestimated radiographic working length[J]. J Endodont,2002,28(2):116- 119.
  • 5Andrej M,Kielbassa, Ursula Muller, et al. Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth[J]. J Oral surgery Oral Medicine Oral pathology,2003,95( 1 ) :94 - 100.
  • 6Kaufman AY,Keila S,Yoshpe M.Accuracy of a new apex locator:an in vitro study[J].Int Endod J,2002,35(2):186-192
  • 7Tamarut T,Kovacevic M,Uhac I.Detection of a transitional ion concentration zone during electronic measurement of root canal length:a study in vitro[J].Int Endod J,2000,33(4):374-380
  • 8Gordon MPJ,Chandler NP.Electronic apex locators[J].Int Endod J,2004,37(7):425-437
  • 9Pratten DH,McDonald NJ.Comparison of radiographic and electronic working lengths[J].J Endod,1996,22(4):173-176
  • 10Kim E,Lee SJ.Electronic apex locator[J].Dent Clin North Am,2004,48(1):35-54

共引文献37

同被引文献4

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部