摘要
本文研制用于微波热疗的频率为91511I,及2450MH_2的两类平板型微带辐射天线该辐射器对人体进行辐射热疗时,与辐射源能很好地匹配,在350MH_4~1000MH_4和2350MH_4~2550MH_2的频率范围,反射系数均小于0.2绝大部分功率可辐射到人体模拟介质中去,辐射深度为3.5cm.
Two Kinds of applicators of 915MH_z and 2450MH_z used in microwave hyperthermia are studied. Both applicators match with the source very well when applied to human model. In the freguency range of 350MH_z~1000MH_z and 2350MH_z~2550MH_z.the reflected coefficients are less than 0.2. Most of the power is transfered to human torso and the depth is 3.5cm.
出处
《山东师范大学学报(自然科学版)》
CAS
1990年第3期26-30,共5页
Journal of Shandong Normal University(Natural Science)
关键词
微带辐射器
微波热疗
癌症
Microwave Hyperthermia
Cancer
Microwave Applicator
Moment Method
Reflected Coefficient