腕式电子血压计血压测量的准确性 被引量:4

Accuracy of Blood Pressure Measurement Using Wrist Blood Pressure Monitor
原文传递
导出
摘要 目的应用腕式血压计测量间接血压与主动脉根部直接测量血压比较,探讨冠状动脉性心脏病(冠心病)患者采用腕式电子血压计测量的准确性。方法将冠脉造影患者(282例,男女比例为181∶101)分为冠心病高血压组86例、单纯冠心病组88例、单纯高血压组57例、正常对照组51例,将测得的腕式血压与中心动脉压通过组内相关系数行一致性评价;以腕式血压与中心动脉压的差值为指标,对高血压、冠心病两因素行析因设计资料的方差分析;同时用英国高血压学会(BHS)仪器设备评级标准对腕式血压计进行准确性评价。结果冠心病对腕式电子血压计收缩压的测量值与主动脉根部直接收缩压的测量值比较,差异有统计学意义(P<0.01);舒张压的测量值差异无统计学意义(P>0.05)。按BHS仪器设备评级标准,冠心病患者收缩压测量评级为D级,但舒张压测量评级为A级。无冠心病的单纯性高血压患者和血压正常患者,对腕式电子血压计收缩压和舒张压的测量值,与主动脉根部直接收缩压与舒张压的测量值的差异均无统计学意义。按BHS仪器设备评级均为A级。结论冠心病对腕式电子血压计收缩压的准确性有影响,而对舒张压无影响。 Objective To observe the accuracy of Wrist Blood Pressure Monitor in patients with Coronary heart disease(CHD).Methods Patients underwent Coronary angiography(282,men/women:181/101)were divided into CHD+hypertension(n=86),CHD(n=88),hypertension(n=57)and the control(n=51).The consistency of indirect wrist and invasive direct central aortic blood pressure were evaluated by the correlation coefficient;the value of difference between indirect wrist and direct aortic pressure was used as index.The accuracy of Wrist Blood Pressure Monitor was evaluated according to the British Hypertension Society(BHS)standard.Results There is significant difference between indirect wrist systolic pressure and direct aortic systolic pressure in patients with CHD(P0.01),however there was no difference of diastolic blood pressure between two methods(P0.05).According to BHS standard,the accuracy level of diastolic blood pressure monitoring was A,meanwhile the accuracy level of systolic blood pressure was only D in the CHD group.There is no significant difference of both systolic and diastolic blood pressure between the wrist and the aortic blood pressure measurement in both hypertension and the control group,accordingly,the accuracy based on BHS level was A.Conclusion Inaccuracy of systolic blood pressure measurement was found in CHD patients using electronic blood pressure monitor.
作者 肖亮 吴延庆
出处 《中华高血压杂志》 CAS CSCD 北大核心 2010年第5期465-468,共4页 Chinese Journal of Hypertension
关键词 冠状动脉性心脏病 血压 电子血压计 腕式 Coronary heart disease Blood pressure Electronic blood pressure monitor wrist
  • 相关文献

参考文献13

  • 1Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals. Part 1: Blood pressure measurement in humans[J]. Circulation, 2005,111 (5) : 697-716.
  • 2Kikuya M, Chonan K, Imai Y, et al. Accuracy and reliability of wrist-cuff devices for selfmeasurement[J]. J Hypertens, 2002,20 (4) :629-638.
  • 3Bernard JG, Eugene B, John DR. Chronic coronary artery disease [A]. In,Eugene B,ed. Heart disease:a text book of cardiovascular medicine[M].5thed. Philadelphia:Saunders, 1997,1289-1365.
  • 4Ahunkan S, Ahunkan E. Validation of the Omron 637IT wrist blood pressure measuring dev-ieewith a position sensor according to the International Protoeolin the elderly[J]. Blood Press Monit, 2006,11(2):97-102.
  • 5李雪红,黄永丽,刘建英.腕式电子血压计用于监测心血管病病人血压的准确性研究[J].护理研究(中旬版),2006,20(9):2416-2417. 被引量:5
  • 6White WB, Berson AS, Robbins C, et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers [J]. Hypertension, 1993, 21 ( 4 ) : 504-509.
  • 7O'Brien E, Petrie J, Litter W, et al. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure mea-suring devices with special reference to ambulatory systems[J]. J Hypertens,1990,8(7) :607-619.
  • 8张瑾,魏菁,席惠芳,何清波,朱鼎良.HEM-6000手腕式血压计的可靠性[J].中华高血压杂志,2007,15(5):407-410. 被引量:9
  • 9苏海.重视血压测量的规范操作[J].高血压杂志,2005,13(10):597-599. 被引量:34
  • 10Bakx C, Oerlemans G, Van den Hoogen H, et al. The influence of cuff size on blood pressure measurement[J]. J Hum Hypertens, 1997,11 (7) : 439-445.

二级参考文献16

  • 1刘必跃.影响电子血压计测量准确的因素[J].计量与测试技术,2004,31(8):24-25. 被引量:31
  • 2郭冀珍,胡大一,李俭春,佟铭,韩雅玲,吴寿岭,何秉贤,王文,曾开淇.电子血压计可靠不可靠?[J].高血压杂志,2005,13(1):5-6. 被引量:21
  • 3王鹏巨,杨原茂.袖带与肢体的力学关系及其对血压测定的影响[J].中华物理医学杂志,1994,16(3):175-178. 被引量:10
  • 4[1]Association for the advancement of medical instrumentation.American National Standard.Electronic or automated sphygmomanometers[J].ANSI/AAMI SP10:2002.
  • 5[2]Kikuya M,Chonan K,Imai Y,et al.Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure[J].J Hypertens,2002,20,629-638.
  • 6[3]Pickering TG,Hall JE,Appel LJ,et al.Recommendations for blood pressure measurement in humans and experimental animals.Part 1:blood pressure measurement in humans.A statement for professionals from the subcommittee of professional and public education of the American Heart Association Council on blood pressure research[J].Circulation,2005,111,697-716.
  • 7[4]Altunkan S,Altunkan E.Validation of the Omron 6371T wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly[J].Blood Press Monit,2006,11,97-102.
  • 8[5]Yarows SA.Comparison on the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled[J].Am J Hypertens,2004,17,54-58.
  • 9Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals part 1: blood pressure measurement in humans[J].Hypertension,2005,45:142-161.
  • 10Marks LA, Groch A. Optimizing cuff width for noninvasive measurement of blood pressure[J].Blood Press Monit,2000,5:153-158.

共引文献44

同被引文献35

引证文献4

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部