增加噪声模拟低剂量胸部CT扫描的研究 被引量:7

Application and research of increasing noise on simulation of the chest scan with low-dose CT
下载PDF
导出
摘要 目的:评估为原始图像添加噪声的模拟低剂量胸部CT的可行性、准确度及其适用范围。方法:为标准剂量下胸部常规CT、薄层高分辨CT各10例和肺结节体模的原始图像添加噪声,模拟常规CT60mA、30mA和薄层高分辨CT160mA、80mA及270mA~10mA(间隔10mA)的低剂量图像。测量主动脉弓水平和肝顶水平主动脉的噪音指数(NI),统计学方法采用配对t检验、Pearson线性相关分析。结果:体模实际扫描图像与模拟图像NI的差异有显著性(P<0.05),呈线性相关;常规CT的NI于主动脉弓层面无差异(P>0.05);于肝顶层面有差异(P<0.05),呈线性相关;薄层高分辨CT在两层面均存在差异(P<0.05),亦存在线性相关。结论:添加噪声取得的模拟图像与实际图像噪声相当,可应用于常规胸部CT平扫的低剂量研究。 Objective: To assess the feasibility, fidelity and applicability simulation of low dose CT scan of chest. Methods: Three sets of raw conventional chest CT scan of ten patients, thin-section high resolution of adding noise on raw data for data were studied, including CT scan of ten patients, and conventional CT scan of lung nodule mannequin. Noise was added to simulate conventional CT scan with tube current of 60 mA and 30 mA, thin-section high resolution CT scan with tube current of 160 mA and 80 mA. For mannequin data, a series of noise were added to simulate tube current from 270 mA to 10mA with interval of 10 mA. Noise index (NI) on levels of aortic arch and liver was measured and compared by paired t test and Pearson correlation analysis. Results: NI was different between scan data and simulated data of mannequin (P〈0.05), which had linear correlation; For conventional CT scan, NI was different on liver level (P〈0.05), while not different on aortic arch level, there had linear correlation; For high resolution CT scan, NI was different on both levels (P〈0.05),which had linear correlation too. Conclusion: For conventional CT scan, the noise between scan data and simulated data is comparable, and adding noise can be used for low dose CT study.
出处 《天津医科大学学报》 2010年第2期329-332,共4页 Journal of Tianjin Medical University
关键词 辐射剂量 模型 理论 体层摄影术 X线计算机 Radiation dose Model Theory Tomography X-ray computed
  • 相关文献

参考文献1

共引文献66

同被引文献42

  • 1孙满顿,滑淑会,李静,安力龙,刘署敏,徐建勇,李刚.低剂量胸部CT扫描技术的临床应用[J].河北职工医学院学报,2005,22(3):11-13. 被引量:4
  • 2Slovis TL.CT and computed radiography: the pictures are great, but is the radiation dose greater than required?[J].Am J Roentgenol,2002,179(1):39-41.
  • 3Entrikin DW,Leipsic JA,Carr JJ.Optimization of radiation dose reduction in cardiac computed tomographic angiography[J]. Cardiol Rev,2011,19(4):163-176.
  • 4American Association of Physicists in Medicine (AAPM) Report 96. The Measurement, Reporting, and Management of Radiation Dose in CT[M]. New York: AAPM, 2008.
  • 5Moscariello A,Takx RA,Schoepf UJ,et al.Coronary CT angiography: image quality,diagnostic accuracy,and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection[J].Eur Radiol,2011,21(10):2130-2138.
  • 6Yu L,Li H,FletcherJG, et al.Automatic selection of tube potential for radiation dose reduction in CT: a general strategy[J]. Med Phys,2010,37(1):234-243.
  • 7Kalender WA,Deak P,Kellermeier M,et al.Applicationand patient size-dependent optimization of x-ray spectra for CT[J]. Med Phys,2009,36 (3):993 - 1007.
  • 8Winklehner A,Goetti R,Baumueller S,et al.Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography:improved doseeffectivenes[J].Invest Radiol,2011,46(12):767-773.
  • 9Leschka S,Stolzmann P,Baumuller S,et al.Performance of dual-energy CT with tin filter technology for the discrimination of renal cysts and enhancing masses[J].Acad Radiol,2010,17(4):526-534.
  • 10Schenzle JC,Sommer WH,Neumaier K,et al.Dualenergy CT of the chest: how about the dose?[J].lnvest Radiol,2010,45 (6) :347 - 353.

引证文献7

二级引证文献82

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部