血细胞分析仪比对试验的应用探讨

Study on the Application of Hematology Analyzer Comparative Test
下载PDF
导出
摘要 目的通过比对实验,了解本实验室不同品牌、型号血细胞分析仪检测结果的准确性和一致性,探讨有效检测范围对一血细胞分析仪的评价。方法选择一台性能较好的血细胞分析仪作为参考比对仪器,采用新鲜抗凝全血及全血质控物同时在几台血细胞分析仪上检测,比较各仪器与参考比对仪器的精密度、高中低值检测结果的相对偏差和有效检测范围。结果MEK-8222K与HMX检测结果间一致性好,偏差在允许误差内。与HMX比较,SF-3000的HCT、WBC测定有效检测范围不合质控要求。结论有效检测范围结合精密度评价血细胞分析仪检测结果的准确性和一致性更全面有效。 Objective Through comparative test, understand the accuracy and conformity of outcomes of different label and style hematology analyzers and explore the evaluation of effective testing range to hematology analyzer. Methods Select one well-functioned hematology analyzer as the referring comparative analyzer and apply fresh anticoagulation whole blood and whole blood quality control product simultaneously on several hematology analyzers to test and compare the precision, relative deviation of high/medium/low testing outcomes and effective testing range of every analyzer and the referring comparative analyzer. Results Testing outcomes of MEK-8222K and HMX had good conformity and deviation was in the allowed error. Compared with HMX, HCT, WBC measuring effective testing range of SF-3000 cannot qualify for quality control. Conclusion Effective testing range combined with accuracy evaluation, the accuracy and conformity of hematology analyzer testing outcomes can be more effective.
作者 王桂东
出处 《中国保健营养(下半月)》 2010年第7期56-57,共2页 China Health Care & Nutrition
关键词 新鲜全血 血细胞分析仪 比对试验 有效检测范围 相对偏差 fresh whole blood hematology analyzer comparative test, effective testing range relative deviation
  • 相关文献

参考文献5

  • 1Takubo T. Tatsumi N, et al. Evaluation of hematological values obtained with reference automated hematology analyzers of six manufacturers [J]. Southeast Asian J Trop Med Public Health. 2002; 33 Suppl 2:62-7.
  • 2胡晓波,李泳,吴驾浦,黄永华,宋颖,许蕾,金大鸣,张锦锋.上海地区血液分析仪准确性的研究和探讨[J].上海医学检验杂志,2001,16(6):355-357. 被引量:22
  • 3Freise K J, Schmidt RL, Gingerich EL, et ah The effect of anticoagulant, storage temperature and dilution on cord blood hematology parameters over time[J]. Int J Lab Hematol, 2008,30(2): 136-138.
  • 4Hedberg P, Lehto T. Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL-DYN sapphire hematology analyzer[J]. Int J Lab Hematol, 2009,31(1):87-95.
  • 5O'Sullivan MB. Quality control of multichannel haematology analysers: critique of current methods and the need for performance goals[J]. Clin Lab haematol, 1990,12(Suppi): 3-12.

共引文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部