摘要
根据重新制作的重铬酸钾(银)剂量计,对γ射线辐射场进行测量。结果表明,剂量计测量重复性较好,相对偏差为2.35%。辐射场的角向和径向剂量呈均匀性分布,而垂直方向在一定高度范围内呈均匀性分布。
The γ-irradiation field is measured by the remade Ag2Cr2O7 - K2Cr2O7 dosimeter. The results show that the dosimeter has high reproducibility with a relative deviation of 2. 35%. The angular and radius direction doses of radiation field are uniformly distributed in the γ-irradiation field. The dose distribution in the perpendicular direction is uniformly distributed in a defined height range.
出处
《实验室科学》
2010年第3期72-75,共4页
Laboratory Science
基金
浙江省教育厅资助项目(项目编号:N20090170)
关键词
辐照装置
重铬酸钾(银)
剂量计
γ辐射场
irradiation equipment
Ag2 Cr2O7 - K2 Cr2O7
dosimeter
γ-irradiation field