摘要
利用ALVIM统计学胸部体模和ROC曲线分析的方法,比较分析了3种显示器(医用、商用、手机)的部分性能,手机显示器的ROC曲线下面积Az值达到0.8048,商用显示器在检测胸部模拟病灶的性能上没有显著性差异。手机显示器能作为便携的终端应用于急诊或床边会诊来浏览医学图像。
ALVIM statistical chest phantom and ROC curve analytic method are used to compare such three kinds of displays as the medical display,commercial display and cellphone display.The ROC Az value of the cellphone display,that is,the area under curve,reaches 0.8048.There is no significant difference when the commercial display is applied to the diagnosis of the simulated chest lesion.The cellphone display can be used in medical image browsing for emergency treatment or bedside consultation.
出处
《医疗卫生装备》
CAS
2010年第6期92-93,95,共3页
Chinese Medical Equipment Journal