酶学参考实验室室间质量评价技术方案的建立 被引量:5

Established Technical Solutions of the Enzymology Reference Laboratory External Quality Assessment
下载PDF
导出
摘要 目的建立酶学参考实验室室间质量评价技术方案,使之符合酶学参考实验室质量要求。方法以γ-谷氨酰氨转移酶(GOT)和乳酸脱氢酶(LDH)为例,按IFCC要求建立该室的酶学参考实验室技术方案,样本的测定采用JCTLM认可的IFCC37℃条件下酶催化活性测量参考方法,以均值和不确定度报告结果。结果返回后,撰写年度室间质量评价总结报告。以酶学参考实验室通过标准评价该室建立的酶学参考实验室室间质量评价技术方案。结果以该室建立的酶学参考实验室室间质量评价技术方案参加GGT和LDH质量评价。检测卫生部临检中心发放的用于监测国家酶学参考实验室的质控样本(GGT样本号是2008101和2008102;LDH样本号是2008103和2008104)。2008年GGT测定结果:该室2008101号样本为:76.27U/L,不确定度1.16U/L;2008102号样本为:225.47U/L,不确定度5.32U/L。国家酶学参考实验室(除掉离群值):2008101号样本为:75.30U/L;2008102号样本为:221.49U/L。2008年LDH测定结果:该室2008103号样本为:225.67U/L,不确定度5.62U/L;2008104号样本为:297.37U/L,不确定度8.94U/L。国家酶学参考实验室(除掉离群值),2008103号样本为:227.90U/L;2008104号样本为:294.03U/L。国家评价标准为均值±5.25%为合格。结论该室建立的酶学参考实验室室间质量评价技术方案符合国家参考实验室质量要求。 Objective To establish technical solutions of the enzymology reference laboratory external quality assessment and to make it consistent with enzyme reference laboratory quality requirements. Methods Example for γ-glutamyl transferase (GGT) and lactate dehydrogenase (LDH), established technical solutions of the enzymology reference laboratory by IFCC reference procedures. Methods of detected samples were approved by IFCC 37℃ and reported the results using mean and uncertainty. The results returned ,wrote an annual external quality assessment summary report. Evaluated enzyme reference laboratories by enzyme reference laboratory external quality assessment technology programs. Results GGT and LDH quality evaluation program. The sample was 2008 external quality assessment:GGT this lab:sample 2008101 was 76.27 U/L,and uncertainty was 1.16 U/L;sample 2008102 was 225.47 U/L and uncertainty was 5.32 U/L. National enzyme reference laboratory (to get rid of outliers) :sample 2008101 was 75.30 U/L;sample 2008102 was 221.49 U/L. LDH this lab:sample 2008103 was 225.67 U/L and uncertainty was 5.62 U/L;sample 2008104 was 297.37 U/L and uncertainty was 8.94 U/L. National enzyme reference laboratory (to get rid of outliers):sample 2008103 was 227.90 U/L;sample 2008104 was 294. 03 U/L. National evaluation standards was mean ±5. 25%. Conclusion This enzyme reference laboratory established EQA technical solutions in line with national reference laboratory quality requirements.
出处 《现代检验医学杂志》 CAS 2010年第3期146-148,共3页 Journal of Modern Laboratory Medicine
基金 国家科技支持计划资金资助项目(2007BA105809).
关键词 参考方法 实验室 参考标准 乳酸脱氢酶 γ-谷氨酰氨转移酶 reference method laboratory reference standards LDH GGT
  • 相关文献

参考文献5

二级参考文献33

  • 1陈文祥.临床检验量值溯源与参考系统[J].中华检验医学杂志,2006,29(1):17-19. 被引量:41
  • 2符颖操,罗茜.实验室间比对结果分析统计方法的探讨[J].理化检验(物理分册),2006,42(6):295-299. 被引量:55
  • 3徐国宾,吴南,王清涛.要重视血清酶学测定的标准化工作[J].临床检验杂志,2007,25(3):161-164. 被引量:25
  • 4Schumann G, Bonom R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 4.Reference procedure for the measurement of catalytic concentration of alanine aminotransferase. Clin Chem Lab Med, 2002, 40:718-724.
  • 5Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 2. Reference procedure for the measurement of catalytic concentration of creatine kinase. Clin Chem Lab Med, 2002, 40:635-642.
  • 6Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 3. Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase. Clin Chem Lab Med, 2002, 40:643-648.
  • 7Schumann G, Aoki R, Ferrero C A, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 8. Reference procedure for the measurement of catalytic concentration of alpha- amylase. Clin Chem Lab Med, 2006, 44-1146-1155.
  • 8International Organization for Standardization ISO18153. In vitro diagnostic medical device-Measurement of quantities in samples of biological origin-Metrological traceability of values for catalytic concentration of enzyme assigned to calibtators and control materials. 2003.
  • 9Schumann G, Bonora R, Cefiotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 2. Reference procedure for the measurement of catalytic activity concentration of creatine kinase. Clin Chem Lab Med, 2002,40: 635-642.
  • 10Schumann G, Bonora R, Cefiotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 3. Reference procedure for the measurement of catalytic activity concentration of lactate dehydrogenase. Clin Chem Lab Med,2002,40: 642-648.

共引文献48

同被引文献24

  • 1徐国宾,吴南,王清涛.要重视血清酶学测定的标准化工作[J].临床检验杂志,2007,25(3):161-164. 被引量:25
  • 2Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the Measurement of catalytic activity concentrations of enzymes at 37℃.Part3.Reference procedure for the measurement of catalytic activity concentration of lactate dehydrogenase[J].Clin Chem Lab Med,2002,40(6):642-648.
  • 3Schumann G, Bonora R, Ceriotti F, et al. IFCC prima- ry reference procedures for the measurement of cata- lytic activity concentrations of enzymes at 37℃. Part6. Reference procedure for the measurement of catalytic activity concentration of 7-g|utamyltrans- ferase[J]. Clin Chem Lab Meal,2002,40(7) :734-738.
  • 4Schumann G,Bonora R,Ceriotti F, et al. IFCC prima- ry reference procedures for the measurement of cata- lytic activity concentrations of enzymes at 37℃. Part3. Reference procedure for the measurement of catalytic activity concentration of lactate dehydrogen- ase[J]. Clin Chem Lab Meal,2002,40(6) : 642-648.
  • 5Schumann G,Bonora R,Ceriotti F. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃.Part6.Reference procedure for the measurement of catalytic activity concentration of γ-glutamyltransferase[J].Clinical Chemistry and Laboratory Medicine,2002,(06):734-738.doi:10.1515/CCLM.2002.126.
  • 6Miller WG, Myers GL . Commutability still matters [J].Clin Chem, 2013,59(9) : 1291-1293.
  • 7Zegers I, Beetham R, Keller T, et al. The importance of commutability of reference materials used as cali- brators: the examples of ceruloplasmin [J]. Clin Chem,2013,59(9) :3122-1329.
  • 8Schumann G, Bonora R, Ceriotti F, et al. IFCC prima- ry reference procedures for the measurement of cata- lytic activity concentrations of enzymes at 37"C. Part3. Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase[J]. Clin Chem Lab Med,2002,40 (16) : 642-648.
  • 9International Organization for Stand ardization(ISO). In vitro diagnostic medical devices-measurement of quantitatias in biological samples-metrologicaI tracea- bility of values assigned to calibrators and control materials[S]. ISO 17511, ISO, 2003.
  • 10Greg Miller W, Myers GL, Lou Gantzer M, et al. Roadmap for harmonization of clinical laboratory measurement procedures[J]. Clin Chem, 2011,57 (8) : 1108-1117.

引证文献5

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部