摘要
This paper presents an approach of active ferromagnetic shimming for C-type permanent magnetic resonance imaging (MRI) magnet. It is designed to reduce inhomogeneity of magnetostatic field of C-type permanent magnet to meet the stringent requirement for MRI applications. An optimal configuration (locations and thicknesses) of active ferromagnetic pieces is generated through calculation according to the initial field map and the demanded final homo- geneity specifications. This approach uses a minimisation technique which makes the sum of squared magnetic moment minimum to restrict the amount of the active ferromagnetic material used and the maximal thickness of pieces stacked at each hole location in the shimming hoards. Simulation and experimental results verify that the method is valid and efficient.
This paper presents an approach of active ferromagnetic shimming for C-type permanent magnetic resonance imaging (MRI) magnet. It is designed to reduce inhomogeneity of magnetostatic field of C-type permanent magnet to meet the stringent requirement for MRI applications. An optimal configuration (locations and thicknesses) of active ferromagnetic pieces is generated through calculation according to the initial field map and the demanded final homo- geneity specifications. This approach uses a minimisation technique which makes the sum of squared magnetic moment minimum to restrict the amount of the active ferromagnetic material used and the maximal thickness of pieces stacked at each hole location in the shimming hoards. Simulation and experimental results verify that the method is valid and efficient.
基金
supported by the National Natural Science Foundation of China (Grant No. 60871001)