期刊文献+

万东i-open磁共振信噪比的测试 被引量:4

Test for SNR of Wandong i-open MRI Equipment
下载PDF
导出
摘要 目的:通过对MRI设备信噪比的检测,及时发现设备存在的问题并加以解决。方法:利用两种测试方法(国家标准方法、包尚联教授的方法)分别测试万东i-open磁共振设备头线圈、颈线圈和体线圈的信噪比,比较不同线圈在两种测试方法下的信噪比并分析结果。结果:测试计算出的信噪比均低于标准值,证实了测试设备的信噪比偏低且不同线圈的信噪比存在差异。结论:所测试MRI设备的信噪比低于标准值,可能与磁屏蔽不良有关。 Object:Through the test for SNR of MRI equipment, discover the problem of equipments and take in to resolve in time. Methods: To test and analyze the SNR of head coil ,neck coil and body coil respectively of Waodong i-open MRI under two kinds of test methods (national standard method and professor Ban Shanglian's method). Results: All the SNR were lower than the Standard value, proved the fact that the SNR of test equipment was lower and verify the differences among the SNR of different coils. Conclusion: The SNR of test MRI equipment below the standard value, maybe related to the magnetic shield problem.
出处 《中国医学物理学杂志》 CSCD 2010年第4期2027-2029,共3页 Chinese Journal of Medical Physics
关键词 磁共振成像 信噪比 线圈 magnmetic resonance imaging SNR coil
  • 相关文献

参考文献5

二级参考文献6

  • 1MansfieldP MorrisPG.生物医学中的核磁共振成像[M].杭州:浙江大学出版社,1987..
  • 2Price RR, Axel L, Morgan T, et al. Quality assurance methods and phantoms for magnetic resonance imaging. Report of AAPM Nuclear Magnetic. Resonance Task Group. No.1. Med Phys, 1990,17:287-295.
  • 3Och JG, Clarke GD, Sobol WT, et al. Acceptance testing of magnetic resonance imaging systems. Report of AAPM Nuclear Magnetic Resonance Task Group. No.6. Med Phys, 1992,19:217-229.
  • 4Kaufman L, Kramer DM, Crooks LE, et al. Measuring signal-to-noise ratios in MR imaging. Radiology, 1989,173:265-267.
  • 5Hashemi RH, Bradley WG. MRI: the basics. New York : Williams, 1997.167-174.
  • 6康立丽,余晓锷.MRI图像信噪比影响因素分析[J].放射学实践,2001,16(6):404-404. 被引量:17

共引文献46

同被引文献33

  • 1张晓槟,金宝荣,苏丽娟.MR图像质量与成像参数的相关性及控制对策[J].数理医药学杂志,2006,19(1):73-75. 被引量:9
  • 2解桂花,徐子森,王雪娜.MRI射频线圈的合理选择与应用[J].青岛大学医学院学报,2006,42(1):84-84. 被引量:2
  • 3李建伟.线圈填充因数与信噪比的关系[J].中华放射学杂志,2007,41(2):131-132. 被引量:4
  • 4杨正汉,冯逢,王霄英,等.磁共振成像技术指南[M].北京:人民军医出版社,2011:233-242.
  • 5郭仁东.光纤通信技术的现状与发展[J].计算机工程应用技术,2008,3(5):1 054-1 055.
  • 6KaufmanL,Kramer D M,Crooks L E,et al.Measuring Signal-toNoise Ratios in MR Imaging[J].Radiology,1989,173:265-267.
  • 7Price R R,Axel L,Morgan T,et al.Quality assurance methods and phantoms for magnetic resonance imaging:report of AAPM Nuclear Magnetic Resonance Task Group No.1[J].Med Phys,1990,17 (2):287-295.
  • 8Och J G,Clarke G D,Sobol W T,et al.Acceptance testing of magnetic resonance imaging systems:report of AAPM Nuclear Magnetic Resonance Task Group No.6[J].Med Phys,1992,19 (1):217-229.
  • 9American College of Radiology. MRI quality control manual. Reston: American College of Radiology, 2004:77-83.
  • 10Och JG, Clarke GD, Sobol WT. Acce77-ptance testing of magnetic resonance imaging systems: Report of AAPM Nuclear Magnetic Resonance Task Group No. 6. Med Phys, 1992,19(1):217-229.