HIFU消融治疗中监控定位的B超伪像 被引量:2

Artifact of type-B ultrasound in monitoring and allocation of HIFU ablation therapy
下载PDF
导出
摘要 目的探讨HIFU消融治疗中B超伪像到探头、探头到目标组织以及治疗头到目标组织三者之间的距离关系。方法在不同位置固定治疗头到目标组织的距离,分别测量伪像到探头以及目标组织到探头之间的距离。结果伪像到探头的距离和目标组织到探头的距离之间呈线性变化关系。结论本实验中伪像的形成主要由B超探头引起,与治疗头的位置基本无关。 Objective To assess the distance relationships among the artifacts of type-B ultrasound and the probe,the probe and the target tissue,as well as the transducer and the target tissue.Methods The distance between the transducer and the target tissue in different locations was located,and the distance between the artifact and the probe,as well as the target tissue and the probe was measured respectively.Results The distance between the artifact and the probe changed linearly when the distance changed between the target tissue and the probe.Conclusion In this experiment,artifacts are mainly caused by ultrasound probe,not correlated much with the position of the transducer.
出处 《中国医学影像技术》 CSCD 北大核心 2010年第6期1165-1168,共4页 Chinese Journal of Medical Imaging Technology
基金 国家自然科学基金重点项目(30830040) 国家自然科学基金面上项目(30970767)
关键词 伪像 超声学 距离 高强度聚焦超声消融 Artifact Ultrasonics Distance High-intensity focused ultrasound ablation
  • 相关文献

参考文献11

二级参考文献55

  • 1邹建中.HIFU高强度聚焦超声治疗肿瘤的超声定位、实时监控及疗效评价[J].临床超声医学杂志,2001,3(S1):23-29. 被引量:15
  • 2李传行,徐国良,黎建军,罗广裕,单宏波,张蓉,李茵.高强度聚焦超声联合化疗治疗恶性肿瘤的初步疗效研究[J].中国微创外科杂志,2004,4(5):395-397. 被引量:7
  • 3张世德,苏玮,薛立军.电路故障诊断中图像定位技术的实现[J].微计算机信息,2005,21(07S):155-156. 被引量:3
  • 4叶慧义,汪伟,陈文治,马林,蔡幼铨,王智彪.MRI对高强度聚焦超声治疗骨肉瘤疗效的随访[J].中华放射学杂志,2005,39(7):726-730. 被引量:5
  • 5吴恩惠.影像诊断学[M].人民卫生出版社,1995.34-35.
  • 6Rabkin BA, Zderic V, Vaezy S. Hyperecho in ultrasound images of HIFU therapy: involvement of cavitation. Ultrasound in Medicine and Biology, 2005,31(7) :947-956.
  • 7Chen WS, Brayman AA, Matula TJ, et al. Inertial cavitation dose and hemolysis produced in vitro with or without Optison. UMB, 2003,29(5) :725-737.
  • 8Khokhlova VA, Bailey MR, Reed JA, et al. Effects o{ nonlinear propagation, cavitation, and boiling in lesion formation by high intensity focused ultrasound in a gel phantom. Journal of the Acoustical Society of America, 2005,119(3) : 1834-1848.
  • 9McLaughlan J, Rivens I, ter Haar G. Cavitation detection in ex vivo bovine liver tissue exposed to high intensity focued ultrasound (HIFU). Biomedical Imaging: Macro to Nano, 4th IEEE International Symposium on Arlington VA, 2007 : 1124-1127.
  • 10Rabkin BA, Zderic V, Vaezy S. HIFU-induced hyperecho in ultrasound images, cavitation activity and thermal behavior. 4th International Symposium on Therapeutic Ultrasound, 2005,754(1) : 43- 46.

共引文献59

同被引文献29

  • 1Kennedy JE. High intensity focused ultrasound in the treatment of solid tumors. Nat Rev Cancer, 2005,5(4):321-327.
  • 2Jolesz FA. MRI-guided foeused ultrasound surgery. Annu Rev Med, 2009(60) :417-430.
  • 3Catane R, Beck A, Inbar Y, et al. MR-guided focused ultrasound surgery (MRgFUS) for the palliation of pain in patients with bone metastases preliminary clinical experience. Ann Oncol, 2007, 18 (1) :163-167.
  • 4Kennedy JE, Ter Haar GR, Cranston D. High intensity focused ultrasound: surgery of the future? Br J Radiol, 2003, 76 (909) : 590-599.
  • 5Furusawa H, Namba K, Thomsen S, et al. Magnetic resonance guided focused ultrasound surgery of breast cancer : reliability and effectiveness. J Am Coll Surg, 2006,203(1) :54-63.
  • 6Wang W, Wang Y, Tang J. Safety and efficacy of high intensity focused ultrasound ablation therapy for adenomyosis. Aead Radb ol, 2009, 16(11) :1416- 1423.
  • 7Jolesz FA, Hynynen K, McDannold N, et al. MR imaging-con- trolled focused ultrasound ablation: a noninvasive image-guided surgery, Magn Reson Imaging Clin N Am, 2005,13(3) :545- 560.
  • 8Fennessy FM, Tempany CM, McDannold N J, et al. Uterine leiomyomas'. MR imaging-guided focused ultrasound surgery-- results of different treatment protocols. Radiology, 2007, 243 (3) :885-893.
  • 9Stewart EA, Gostout 13, Rabinovici J, et al. Sustained relief of leiomyoma symptoms by using focused ultrasound surgery. Ob- stet Gynecol, 2007,110(2 Pt 1) :279-287.
  • 10Kennedy JE, Terhaar GR, Wu F, et al. Contrast-enhanced ul-trasound assessment to response to high-intensity focused ultra- sound. Ultrasound Med Biol, 2004,30(6) :851-854.

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部