自动管电流调节技术在胸部低剂量CT筛查中的临床价值 被引量:53

Clinical value of automatic tube current modulation in chest screening with low dose CT
下载PDF
导出
摘要 目的探讨16层螺旋CT自动管电流调节(ATCM)技术在胸部低剂量CT筛查中的临床价值。方法 94名常规体检人员分为两组:A组45名,扫描方案为固定噪声指数(NI)为14,应用ATCM技术自动调节管电流(20~200mA);B组49名,扫描方案为固定管电流40mAs。两名医师采用双盲法对肺尖、中肺、下肺图像进行图像质量评分(IQS),测量图像标准差(SD)值,记录加权CT剂量指数(CTDIw),并进行统计学分析。结果肺尖、中肺、下肺SD值:A组为13.62±2.07、11.70±0.85、11.89±1.26,B组为17.20±2.66、11.29±2.32、12.82±2.49;IQS:A组为1.82±0.36、1.58±0.38、1.66±0.34,B组为2.30±0.47、1.67±0.42、1.74±0.48。在肺尖,A组的SD值和IQS低于B组(P<0.05)。A、B两组CTDIw比较:受检者体质量指数(BMI)≤27kg/m2时,A、B两组差异无统计学意义;BMI>27kg/m2时,A组高于B组(P<0.05)。结论胸部低剂量CT筛查中,应用ATCM技术可在更大的密度衰减范围和体质量较大的人群中得到更高质量的图像。 Objective To investigate the clinical value of automatic tube current modulation(ATCM)in low dose chest scaning with 16-slice spiral CT.Methods Ninety-four healthy subjects for routine health examination were enrolled in this study and divided into two groups.Forty-five subjects of group A underwent chest CT scanning using ATCM(20—200 mA)with noise index(NI)of 14,and 49 subjects of group B underwent chest CT scanning with a fixed tube current of 40 mAs.CT images were analyzed by two radiologists with double blind method.Image quality scores(IQS)of lung apex,middle lung and lower lung were made,the image standard deviation(SD)was measured,and the weighted CT dose index(CTDIw)was also recorded.The statistical analysis was performed.Results SD values in lung apex,middle and lower lungs was 13.62±2.07,11.70±0.85,11.89±1.26 in group A,and 17.20±2.66,11.29±2.32,12.82±2.49 in group B,as well as the corresponding IQS of these three sites were 1.82±0.36,1.58±0.38,1.66±0.34 in group A,and 2.30±0.47,1.67±0.42,1.74±0.48 in group B,respectively.SD value and IQS of group A were lower than those of group B in lung apex(P〈0.05).For the subjects with body mass index(BMI)≤27 kg/m^2,there was no statistical difference of CTDIw between the two groups(P〉0.05).When BMI27 kg/m^2,CTDIw of group A was higher than that of group B(P〈0.05).Conclusion The application of ATCM in low dose chest CT screening could provide more homogeneous images with better image quality and more attenuation areas in obese subjects.
出处 《中国医学影像技术》 CSCD 北大核心 2010年第6期1169-1172,共4页 Chinese Journal of Medical Imaging Technology
关键词 低剂量 体层摄影术 X线计算机 噪声指数 自动管电流调节 图像质量 Low-dose Tomography X-ray computed Noise index Automatic tube current modulation Image quality
  • 相关文献

参考文献14

  • 1Roberts HC,Patsios D,Paul NS,et al.Lung cancer screening with low-dose computed tomography:Canadian experience.Can Assoc Radiol J,2007,58(4):225-235.
  • 2Itoh S,Ikeda M,Mori Y,et al.Lung:feasibility of a method for changing tube current during low-dose helical CT.Radiology,2002,224(3):905-912.
  • 3Rampado O,Marchisio F,Izzo A,et al.Effective dose and image quality evaluations of an automatic CT tube current modulation system with an anthropomorphic phantom.Eur J Radiol,2009,72(1):181-187.
  • 4Papadakis A,Perisinakis K,Damilakis G.Angular on-line tube current modulation in multidetector CT examinations of children and adults:the influence of different scanning parameters on dose reduction.Med Phys,2007,34(7):2864-2874.
  • 5Menke J.Comparison of different body size parameters for individual dose adaptation in body CT of adults.Radiology,2005,236(2):565-571.
  • 6Kalra MK,Maher MM,Toth TL,et al.Techniques and applications of automatic tube current modulation for CT.Radiology,2004,233(3):649-657.
  • 7Bach PB,Silvestri GA,Hanger M,et al.Screening for lung cancer:ACCP evidence-based clinical practice guidelines (2nd ed).Chest,2007,132(3 Suppl):69S-77S.
  • 8韦进,王小文,吴广仕,麦兆绵,揭育添.多排螺旋CT非螺旋低剂量扫描筛查肺结节[J].中国介入影像与治疗学,2009,6(6):563-566. 被引量:18
  • 9Swensen SJ,Jett JR,Hartman TE,et al.CT screening for lung cancer:five-year prospective experience.Radiology,2005,235(1):259-265.
  • 10New York Early Lung Cancer Action Project Investigators.CT Screening for lung cancer:diagnoses resulting from the New York Early Lung Cancer Action Project.Radiology,2007,243(1):239-249.

二级参考文献18

  • 1Maikitaro R, Paakko P, Huhti E, et al. Prospective populationbased study on the survival of patients with lung cancer. Eur Respir J, 2002, 19(6) :1087-1092.
  • 2Greenlee RT, Murray T, Bolden S, et al. Cancer statistics, 2000. CA Cancer J Clin, 2000,50(1) :7-33.
  • 3Padilla J, Calvo V, Penalver JC, et al. Survival and risk model for stage I B non small cell lung caner. Lung Cancer, 2002, 36 (1) :43-48.
  • 4Monnier-Cholley L, Arrive L, Porcel A, et al. Characteristics of missed lung cancer on chest radiographs: a French experience. Eur Radiol, 2001,11(4) :597-605.
  • 5Li F, Sone S, Abe H, et al. Lung cancers missed at lowdose hell cal CT screening in a general population: comparison of clinical histopathologic and imaging findings. Radiology, 2002, 225(3) 673-683.
  • 6Ohno Y, Takenake D, Watanabe H, et al. Low-dose lung cancer screening using multidetector-row CT system: utility of the MIP image for improvement of detectability. Radiology, 2001, 221 (Suppl) :313.
  • 7Starck G, Lonn L, Cederblad A, et al. A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol, 2002,75(890) : 140-150.
  • 8Diederich S, Wormanns D, Semik M, et al. Screening for early lung cancer with low dose spiral CT: prevalence in 817 asymptomatic smokers. Radiology, 2002,222(3) : 773-781.
  • 9Swensen SJ, Jett JR, Hartman TE, et al. Lung cancer screen ing with CT: Mayo Clinic experience. Radiology, 2003,226 (3) : 756-761.
  • 10Nawa T, Nakagawa T, Kusano S, et al. Lung cancer screening using low-dose spiral CT. Chest, 2002,122(1) :15-20.

共引文献39

同被引文献458

引证文献53

二级引证文献408

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部