加速器锥形束CT成像参数对空间分辨率的影响 被引量:10

Variations of spatial resolution with imaging parameters for kilovoltage cone-beam CT
下载PDF
导出
摘要 目的分析加速器机载kV锥形束CT(CBCT)成像参数对空间分辨率的影响。方法采用CatPhan600模体的CTP528模块,分别改变管电压、管电流、准直器、扫描范围和重建精度,测量XVICBCT的空间分辨率。结果管电压从80kV增至140kV时,空间分辨率从5lp/cm增加到7lp/cm。管电流对空间分辨率影响较小。准直器型号、重建精度和机架旋转范围对空间分辨率影响较大。结论 CBCT的影像质量随成像条件变化范围很大,实际应用时需要根据条件合理选择参数,减少患者辐射剂量的同时得到最佳的影像质量。 Objective To evaluate the impact of preset imaging parameters on spatial resolution of Linac-integrated kV cone-beam CT(CBCT)unit.Methods The spatial resolution in CBCT mode was evaluated with CPT 528 module of CatPhan 600 phantom with different combinations of kV,mAs,collimator,scanning ranges and reconstruction resolution.Results The spatial resolution increased from 5 lp/cm to 7 lp/cm when the tube voltage varied from 80 kV to 140 kV.Current and exposure time per frame had little influence on the spatial resolution.The spatial resolution mainly depended on collimator,reconstruction resolution,and gantry rotation range.Conclusion The image quality of CBCT varies with the preset scanning parameters.Parameters should be reasonably chosen according to the patient's anatomy,so as to compromise image quality and radiation dose.
出处 《中国医学影像技术》 CSCD 北大核心 2010年第6期1173-1175,共3页 Chinese Journal of Medical Imaging Technology
关键词 影像引导放射治疗 锥形束体层摄影术 空间分辨率 Image-guided radiotherapy Cone-beam computed tomography Spatial resolution
  • 相关文献

参考文献10

二级参考文献36

  • 1于金明,袁双虎.图像引导放射治疗研究及其发展[J].中华肿瘤杂志,2006,28(2):81-83. 被引量:99
  • 2戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 3徐晓东,邵军明,孔军.西门子多排螺旋CT新技术[J].CT理论与应用研究(中英文),2006,15(2):22-27. 被引量:15
  • 4胡逸民 张红志 戴建荣 胡逸民 主编.治疗计划的设计与执行[A].胡逸民,主编.肿瘤放射物理学[C].北京:原子能出版社,1999.463-471.
  • 5S.Schaller, K.Stierstorfer, H.Bruder and etc.Novel approxinate approach for high-quality image reconstruction in helical cone beam CT at arbitrary pitch[J]. Proceedings of SPIE,2001,Vol.4322:113-127.
  • 6T.Flohr, K.Stierstorfer and etc. Image reconstruction and image quality evaluation for a 16-slice CT scanner[J]. Medical Physics, April 2003. Vo130,No.5:832-845.
  • 7T.Flohr, H.Bruder and etc..New technical developments in multi-slice CT: partl: approaching isotropic resolution with sub-millimeter 16-slice scanning[J]. Fortschr Rontgenstr, 2002,(174):839-.845.
  • 8M. Kachelriess, M. Knaup, C. Penssel, and W. Kale, nder.Flying Focal Spot _FFS_ in Cone-Beam CT[J]. Records of the 2004 IEEE Medical Imaging Conference, 2005, pp. 3759-3763.
  • 9J.Hsieh, R.C.Molthen, C.A.Dawson, and R.H.Johnson.An iterative approach to the beam hardening correction in cone beam CT[J]. Med.Phys.27(1),pp.23-29(2000).
  • 10H.Vestner and H.Wallschlager.Basic concepts in computed tomography[C].29,Jun.,2005. Siemens Forchheim.

共引文献71

同被引文献82

引证文献10

二级引证文献30

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部